Supports 4- and 6-wire load cell connections legal seal is affixed to the back of the indicator, • Two configurable digital inputs this is the only mode in which the IQ plus 350 can • Electronic data processing (EDP) port for full operate.
Entry Mode Nicht Geeicht Front Panel Keypad Entry The indicator is not in normal mode. The Figure 1-1 shows the IQ plus 350 front panel keys and Mode Nicht Geeicht ) annunciator is shown the key functions assigned in normal mode.
Indicator Operations 1.4.4 Acquire Tare 1. Place container on scale and wait for the Basic IQ plus 350 operations are summarized below: standstill annunciator ( 1.4.1 Toggle Gross/Net Mode 2. Press the key to acquire the tare weight TARE Press the...
LOCATION OF INSTALLED ANALOG OUTPUT MODULE SERIAL COMM & DIGITAL IN Cable tie mounts on inside of enclosure top Figure 2-2. IQ plus 350 CPU and Power Supply Board, Showing Installed Cables and Cable Ties 2.3.1 Load Cells To attach cable from a load cell or junction box,...
SETUP SWITCH LOCATION OF INSTALLED ANALOG OUTPUT MODULE –EXC +EXC SHIELD (ANALOG OUT) –SENSE +SENSE –SIG +SIG PWR2 PWR1 KEYPAD CONNECTOR BROWN BLUE WIRE WIRE Figure 2-3. IQ plus 350 CPU and Power Supply Board IQ plus 350 Installation Manual...
2.3.3 Analog Output If the optional analog output module is installed, If you must remove the IQ plus 350 CPU board, use attach the output cable to connector J1 on the analog the following procedure: output board. Table 2-3 lists the analog output pin 1.
Replacement Parts Table 2-4 lists replacement parts for the IQ plus 350, including all parts referenced in Figures 2-5 through 2-9. Number Description (Quantity) Figure 14626 Kep nuts, 8-32NC hex (7) Figure 2-6 on page 10 41300 Display and CPU board assembly, 115 VAC (1)
Page 13
To Line Filter 20/4X/B 24/4X See Line Filter Ground Post Assembly Figure 2-5. IQ plus 350 Backplate and Tilt Stand Assemblies Installation...
Page 14
1/9X/A AC power in from line filter To setup switch 30/8X 3/4X Figure 2-6. IQ plus 350 Enclosure and CPU Board To backplate ground post Ground Post Assembly To line filter Setup switch connection to JP4 on CPU board To power cord AC input to CPU board JP7 Figure 2-7.
Page 15
9/10X/D 2 5 / 6 X Figure 2-9. IQ plus 350 Enclosure, Showing Bezel, Overlay, and Backplate Gasket Installation...
Configuration Methods mode. The IQ plus 350 indicator can be configured by using 4. Start the Revolution Scaleware program. the front panel keys to navigate through a series of Figure 3-1 shows an example of one of the Revolution configuration menus or by sending commands or...
Front Panel Configuration The IQ plus 350 indicator can be configured using a series of menus accessed through the indicator front panel when the indicator is in setup mode. Table 3-1 summarizes the functions of each of the main menus.
(see Figure 3-4). Menu Structures and Parameter Descriptions The following sections provide graphic representations of the IQ plus 350 menu structures. In the actual menu structure, the settings you choose under each parameter are arranged horizontally. To save page space, menu choices are shown in vertical columns.
3.2.2 Format Menu CONFIG XXXXXXX FORMAT CALIBR SERIAL PROGRM XXXXXXX PFORMT DIG IN XXXXXXX ALGOUT XXXXXXX VERS PRIMAR SECNDR DSPRAT DECPNT DSPDIV UNITS DECPNT DSPDIV UNITS MULT MULEXP 250MS 500MS decimal 750MS 888888 88888.8 0.453592 position 888880 number 1500MS 888888 888880 2SEC 8.88888...
Page 22
To toggle between primary and secondary units, press the UNITS key. MULEXP decimal Multiplier exponent. Sets the decimal position for multiplier values of 10 or more. position Table 3-3. Format Menu Parameters (Continued) IQ plus 350 Installation Manual...
3.2.3 Calibration Menu See Section 4.0 on page 25 for calibration procedures. CONFIG XXXXXXX FORMAT CALIBR SERIAL PROGRM XXXXXXX PFORMT DIG IN XXXXXXX ALGOUT XXXXXXX VERS WZERO WVAL WSPAN REZERO *CAL* CAL* *CAL* Display and edit Display and edit Press Enter to Display and edit test weight value span calibration...
3.2.4 Serial Menu See Section 7.3 on page 35 for information about the IQ plus 350 serial data format. CONFIG FORMAT XXXXXXX CALIBR SERIAL PROGRM PFORMT XXXXXXX DIG IN ALGOUT XXXXXXX XXXXXXX VERS PRINT STREAM PRNDES BAUD BITS TERMIN EOLDLY...
SERIAL Menu Parameter Choices Description EOLDLY End-of-line delay. Sets the delay period, in 0.1-second intervals, from when a formatted line is number terminated to the beginning of the next formatted serial output. Value specified must be in the range 0-255, in tenths of a second (10 = 1 second). Table 3-5.
Scroll right in format string characters of format Display and edit active character and Decrement ASCII value of active character Increment ASCII value of active character ASCII value Delete active character Figure 3-10. Print Format Menu IQ plus 350 Installation Manual...
3.2.7 Digital Input Menu CONFIG FORMAT CALIBR SERIAL PROGRM PFORMT DIG IN ALGOUT VERS XXXXXXX XXXXXXX XXXXXXX XXXXXXX DIGIN1 DIGIN2 ZERO ZERO TARE TARE NT/GRS NT/GRS UNITS UNITS DSPTAR DSPTAR PRINT PRINT CLRCN CLRCN KBDLOC KBDLOC HOLD HOLD Figure 3-11. Digital Input Menu DIG IN Menu Parameter Choices...
Version menu: when selected, the indicator displays the installed software version number. CONFIG FORMAT XXXXXXX CALIBR SERIAL PROGRM PFORMT XXXXXXX DIG IN ALGOUT XXXXXXX XXXXXXX VERS Software version Figure 3-13. Version Menu IQ plus 350 Installation Manual...
Calibration ™ The IQ plus 350 can be calibrated using the front panel, EDP commands, or the Revolution Scaleware configuration utility. Each method consists of the following steps: • Zero calibration • Entering the test weight value • Span calibration •...
Settings display are filled in. Click to save the new Exit values and return to the Revolution Scaleware main menu; to restore the previous calibration values, click Restore Settings Figure 4-3. Revolution Scaleware Calibration Display IQ plus 350 Installation Manual...
EDP Commands The IQ plus 350 indicator can be controlled by a personal computer or remote keyboard connected to Command Function the indicator EDP port. Control is provided by a set of EDP commands that can simulate front panel key...
List all parameter values 5.1.4 Parameter Setting Commands VERSION Write IQ plus 350 software version Parameter setting commands allow you to display or change the current value for a particular configuration Write current displayed weight with units parameter (Tables 5-3 through 5-10).
Page 33
Command Description Values PRI.DECPNT Primary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888, 888880 PRI.DSPDIV Primary units display divisions 1D, 2D, 5D PRI.UNITS Primary units LB, KG, OZ, TN, T, G, NONE SEC.DECPNT Secondary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888, 888880 SEC.DSPDIV Secondary units display divisions 1D, 2D, 5D...
Transmit net weight in non-displayed units Transmit tare weight in non-displayed units Query system error conditions nnnnn nnnnn See Section 7.1.2 on page 34 for detailed information about the XE command response format. Table 5-11. Normal Mode EDP Commands IQ plus 350 Installation Manual...
SERIAL menu match the baud rate, bits, and parity IQ plus 350 and the attached scale is not changed, settings configured for the serial port on the PC. Set recalibration is not required.
KPRINT EDP command is received. If the indicator is <N>, and <T>). in gross mode, the GFMT print format is used; if in The default IQ plus 350 print formats are shown in net mode, NFMT is used. Table 6-2:...
IQ plus 350 front panel SMALLTOWN (see the ASCII character chart on page 36) and are shown as blanks. The IQ plus 350 can send or receive 1345 LB GROSS any ASCII character; the character printed depends on the particular ASCII character set implemented for the receiving device.
Error Messages The IQ plus 350 indicator provides a number of error messages. When an error occurs, the message is shown on the indicator LCD display. Error conditions can also be checked remotely by using the XE EDP command as described in Section 7.1.2.
Continuous Output (Stream) Format Figure 7-1 shows the continuous output format sent to the IQ plus 350 EDP or printer port when the STREAM parameter (SERIAL menu) is set to either EDP or PRN. <STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>...
Characters that strings on the IQ plus 350 PFORMT menu. The actual can be shown on the IQ plus 350 display are indicated character printed depends on the character mapping by shading in the decimal value columns in Table 7-4.
Page 41
ASCII ASCII ASCII ASCII Ç á α ü í β Γ é ó π â ú ä ñ Σ à Ñ σ å ª µ τ ç º ê ¿ Φ ë Θ è ¬ Ω δ ï î ∞ ì...
Conversion Factors for Secondary Units Primary Unit x Multiplier Secondary Unit The IQ plus 350 has the capability to mathematically grains 0.064799 grams convert a weight into many different types of units and 0.002286 ounces instantly display those results with a press of the key.
Digital Filtering The IQ plus 350 uses averaged digital filtering to reduce the effect of vibration on weight readings. Adjustable threshold and sensitivity functions allow quick settling by suspending filter averaging, allowing the weight reading to jump to the new value. Figure 7-2 shows the digital filter parameters on the CONFIG menu.
3 and 4 to re-adjust the zero value as required. • For current output, connect ammeter 6. Return to normal mode. Analog output leads to pins 1 and 2 function can be verified using test weights. IQ plus 350 Installation Manual...
Test Mode To enter test mode, press and hold the setup switch until the front panel display shows the word TEST In addition to normal and setup modes, test mode After about three seconds, the test mode display provides a number of diagnostic functions for the IQ automatically shifts to the first test menu function, A/ plus 350, including: DTST.
Page 46
Press and hold Enter key to send ASCII “U” characters (decimal 85) from the serial port. ECHO R Echo received characters Press and hold Enter key to view characters received at serial port. IQ plus 350 display shows lower-case characters as blanks. NOTE: Table 7-7. Test Menu Functions IQ plus 350 Installation Manual...
Specifications Digital Specifications Microcomputer Hitachi H8/3834 main processor @ Power 9.8304 MHz Line Voltages 115 or 230 VAC Digital Inputs 2 inputs, TTL or switch closure, active-low Frequency 50 or 60 Hz Digital Filters 3 filters, software selectable Power Consumption 66 mA @ 115 VAC (8W) 33 mA @ 230 VAC (8W) Serial Communications...
IQ plus 350 Limited Warranty Rice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed by the Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects in materials and workmanship for two years.