Changing the Initial Brightness of Dia-illumination LED

13.2
The brightness of the dia-illumination LED is saved for each address of the nosepiece, and the brightness previously set
for the address is automatically set if the nosepiece is switched. Note that the brightness is reset once the illumination is
turned off, and is set to the configured initial brightness upon next use. The initial brightness can be changed as follows:
(1)
Check that the setting mode switch is set to the left side
(operation mode).
(2)
Bring each of the nosepiece addresses into the optical
path and adjust the brightness for all addresses using
the dia-illumination LED brightness control knob.
(3)
(If the toggle function is enabled) Press the toggle
pattern switch until the toggle function is turned OFF.
(4)
Set the setting mode switch to the right side to switch to
the setting mode.
(5)
Hold down the capture button on the main body, and set
the setting mode switch to the left side to go back to the
operation mode.
Nosepiece address buttons 1 to 6 light for a few seconds to
indicate that setup is complete.
For each of the nosepiece addresses, the brightness set in
step (2) is set as the initial brightness.