Bring a desired objective into the optical path.
Press the nosepiece forward/reverse rotation buttons on the main body, or the nosepiece address button on the
remote control pad to move the desired objective into the optical path.
When using an oil immersion type objective, apply immersion oil between the specimen and the objective.
(→Chapter 2 "14 Oil Immersion" for details)
8
Focus on the specimen.
(1)
Look into the eyepiece, and adjust the brightness of
the field of view with the ND filter of the
epi-fluorescence attachment.
(2)
Rotate the stage knob to move the stage and bring
the target into the optical path.
(3)
If the specimen is not in focus, turn the focus knob to
focus on it.
9
Adjust the field diaphragm of the epi-fluorescence attachment.
Use the field diaphragm lever of the epi-fluorescence
attachment to adjust the field diaphragm so that it almost
circumscribes the field of view.
Chap. 1-5
Chapter 1
Non-fluorescent immersion oil
Use Nikon designated non-fluorescent immersion oil.
Size of the field diaphragm
Normally, adjust the field diaphragm so that it
almost circumscribes the field of view. Excessively
opening the field diaphragm will result in stray light
entering the field of view, generating flare and
reducing the image contrast. In addition, the
specimen will become decolorized over a wider
area.
Adjustment timing for the field
diaphragm
Be sure to adjust the field diaphragm each time
you change the objective.
Microscopy Procedures
Brightness adjustments with ND filters
Field Diaphragm
Circumscribe around the field of view
Adjusting the field diaphragm
36
Field
diaphragm
lever
CUBE
1
1
2
3
4
Field of view