Physical Properties - Nikon Eclipse Ci-E Instruction Manual

Upright motorized microscope
Hide thumbs Also See for Eclipse Ci-E:
Table of Contents

Advertisement

Power cord
CI-FL Epi-fluorescence Attachment for Nikon Microscope
Model
Optical system
Fluorescent turret
Field diaphragm
ND filter
Shutter
Supported illuminator
3

Physical Properties

ECLIPSE Ci-E Nikon Microscope
Model
Operating conditions
Transport/storage
conditions
External dimensions and
mass (Main body only)
Chap. 6
Chapter 6
Specifications and Safety Standards
 When used in 100-120 V region, outside Japan
UL listed detachable power cord set, 3 conductor grounding
(3 conductor grounding Type SVT, No.18 AWG, 3 m long maximum, rated at 125 VAC
minimum)
 When used in 220-240 V region
Detachable power cord set approved according to EU/EN standard, 3 conductor
grounding
(3 conductor grounding Type H05VV-F 1 mm
minimum)
 When used inside Japan
PSE approved detachable power cord set, 3 conductor grounding
(3 conductor grounding Type VCTF 3 x 0.75 mm
minimum)
CI-FL Epi-fluorescence attachment
Infinity-corrected CF optical system
Variable intermediate magnification: 1x
Manual quadruple turret
Manual adjustment
Manual, 3 filters (slider type; ND4, ND8, ND16)
Manual (using a lever at the bottom of the turret)
HG precentered fiber illuminator
ECLIPSE Ci-E
Temperature:
Humidity:
Altitude:
Pollution degree:
Overvoltage category:
Electrical shock protection class:
Indoor use only
Temperature:
Humidity:
External dimensions:
Mass (reference):
2
, 3 m long maximum, rated at 250 VAC
2
, 3 m long maximum, rated at 125 VAC
0°C to +40°C
60% RH max. (no condensation)
2000 m max.
Degree 2
Category II
Class I
-20°C to +60°C
90% RH max. (no condensation)
223 (W) x 331.5 (H) x 331 (D) mm
(excluding projections)
Approx. 10 kg
90

Advertisement

Table of Contents
loading

Table of Contents