Electrical System; General - Nikon Eclipse Ci-E Instruction Manual

Upright motorized microscope
Hide thumbs Also See for Eclipse Ci-E:
Table of Contents

Advertisement

2

Electrical System

General

2.1
Power supply
Problem
There is no power even
though the power switch
is on.
Illumination
Problem
The dia-illumination LED
is not illuminated.
The dia-illumination LED
cannot be turned
ON/OFF with the
ON/OFF switch.
A brightness cannot be
adjusted with the
dia-illumination LED
brightness control knob.
Motorized nosepiece
Problem
The motorized
nosepiece does not
move.
The motorized
nosepiece skips
addresses.
Motorized swing-out condenser
Problem
The motorized swing-out
condenser does not
move.
The motorized swing-out
condenser's top lens is
not in the designated
position.
Operation of the
motorized swing-out
condenser is not linked
to the rotation of the
nosepiece.
Chapter 4
Cause
The power cord is not connected, or is
connected improperly.
Cause
There is no power supplied.
The microscope is set to the setting mode.
A control is given to DS-L4 or Ni Setup
Tool. Otherwise, an operation has been
disabled due to DS-L4 or Ni Setup Tool
settings.
A control is given to DS-L4 or Ni Setup
Tool. Otherwise, an operation has been
disabled due to DS-L4 or Ni Setup Tool
settings.
Cause
The microscope is set to the setting mode.
An operation has been disabled due to
DS-L4 or Ni Setup Tool settings.
The toggle function is ON.
Cause
The motorized swing-out condenser cable
is not connected.
The top lens has been misaligned due to
external force.
Interlocking is not configured.
Troubleshooting
Connect the cable properly.
(→Chapter 3, "2 Assembly for Bright-field
Microscopy - 9 Connect the power cord".)
Plug in the power cord.
(→Chapter 3, "2 Assembly for Bright-field
Microscopy - 9 Connect the power cord".)
Set the setting mode switch on the remote
control pad to the left side to switch to the
operation mode.
Change the setting with DS-L4 or Ni Setup
Tool.
Change the setting with DS-L4 or Ni Setup
Tool.
Set the setting mode switch on the remote
control pad to the left side to switch to the
operation mode.
Change the setting with DS-L4 or Ni Setup
Tool.
Press the toggle pattern switch on the remote
control pad until the toggle function is turned
OFF.
(→Chapter 2, "12.2 Switching between Two
Objectives (Toggle)".)
Connect the cable properly.
(→Chapter 3, "2 Assembly for Bright-field
Microscopy".)
Turn the power OFF, and then turn it ON
again.
Configure interlocking.
(→Chapter 2, "13.1 Configuring the Motorized
Swing-out Condenser".)
85
Measure
Measure
Measure
Measure
Chap. 4