Attach The Lamp; Attach The Dia-Illumination Lamphouse - Nikon Eclipse Ni-E Instruction Manual

Upright motorized microscope
Hide thumbs Also See for Eclipse Ni-E:
Table of Contents

Advertisement

6

Attach the lamp.

Tool: Hex wrench (3 mm across flats)
Lamp handling precautions
Avoid touching the glass surface of the lamp with
your bare hands.
(1)
Loosen a lamphouse-cover fixing screw and lift up
the cover to remove it.
(2)
Push in the lamp clamp lever to open the socket pin
hole. Attach the lamp while holding down the lever.
Put the lamp clamp lever back to its original position.
Specified lamp: PHILIPS7724
or OSRAM HLX64623
(3)
Reattach the cover back to its original position and
tighten the lamphouse-cover fixing screw.
7

Attach the dia-illumination lamphouse.

Tool: Hex driver (2 mm across flats)
(1)
Insert the lamphouse at the rear of the microscope
and tighten the fixing screws.
(2)
Connect the dia-illumination lamphouse cable to the
lamp connector of the control box A using a relay
cable.
Chapter 1
Assembly
Fixing screw
Lamphouse
fixing screw
Securing the dia-illumination lamphouse
9
Insert the hex wrench and
loosen the fixing screw.
Removing the lamphouse cover
Lamp
Attaching the lamp
Lamphouse
fixing screw
ND
ND
NC B
8
32
11
OU T
IN
Lamphouse
cover
Lamp
clamp lever

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents