Nikon Eclipse Ci-E Instruction Manual page 21

Upright motorized microscope
Hide thumbs Also See for Eclipse Ci-E:
Table of Contents

Advertisement

Microscopy operation
13
Select the desired objective.
Press the nosepiece forward/reverse rotation buttons, or
the nosepiece address button on the remote control pad to
move the desired objective into the optical path.
1-100x swing-out condenser
Mounting a 1-100x swing-out condenser on the
Ci-E main body when using the 1x objective may
result in uneven illumination around the field of
view.
Toggle function
Use the toggle function for easy switching
between two objectives. (→Chapter 2, Section
12.2 for details)
14
Adjust the aperture diaphragm.
Turn the aperture diaphragm lever on the condenser to
adjust the aperture diaphragm so that it is set to 70 to 80%
of the numerical aperture of the objective used.
Adjustment timing for the aperture
diaphragm
Be sure to adjust the aperture diaphragm each
time you change the objective.
(You can see the aperture diaphragm image with
the centering telescope.)
15
Focus on the specimen.
(1)
Look into the eyepiece, and adjust the brightness of
the field of view by turning the dia-illumination
brightness control knob.
(2)
Rotate the stage knob to move the stage and bring
the target into the optical path.
(3)
If the specimen is not in focus, turn the focus knob to
focus on it.
Chapter 1
Microscopy Procedures
(→See Chapter 2, "6 Adjusting the Aperture Diaphragm" for details)
9
OFF
1
2
TOGGLE
PATTERN
MEMORY
Bringing a desired objective into the optical path
Objective
pupil
Right size of the aperture diaphragm
Chap. 1-1
Aperture diaphragm
image
70 to 80
100