Hide thumbs Also See for Eclipse Ci-E:
Table of Contents

Advertisement

3
Set the setting mode switch to the left side (operation mode).
4
Turn on the dia-illumination LED.
5
Lower the condenser slightly from the uppermost position.
6
Fully open the field diaphragm and aperture diaphragm.
7
Bring the 10x objective into the optical path.
8
Place a specimen on the stage, and move the stage to bring the target into view.
9
Remove the analyzer and the polarizer from the optical path.
Pull out the analyzer IN/OUT knob from the analyzer tube
for simple polarization to remove the analyzer from the
optical path.
The polarizer unit for simple polarization has not yet been
installed at this point.
10
Focus on the specimen.
details)
11
Adjust the diopter.
12
Adjust the interpupillary distance.
13
Focus and center the condenser.
Chapter 1
Microscopy Procedures
(→See Chapter 2 "2 Focusing on the Specimen (Vertical Stage Movement)" for
(→See Chapter 2, "4 Adjusting the Diopter" for details)
(→See Chapter 2, "5 Focusing and Centering the Condenser" for details)
Removing the analyzer from the optical path