Attach The Dia Motorized Shutter (Optional); Attach The Lamp; Attach The Lamphouse For Dia-Illumination - Nikon Eclipse Ni-U Instruction Manual

Upright microscope assembly/maintenance
Table of Contents

Advertisement

2

Attach the DIA motorized shutter (optional).

The motorized shutter is attached by Nikon.
Contact your nearest Nikon representative when the DIA motorized shutter needs to be attached or removed.
3

Attach the lamp.

Tool: Hex wrench (3 mm across flats)
Lamp handling precautions
Avoid touching the glass surface of the lamp with
your bare hands.
(1)
Loosen the lamphouse cover fixing screw and lift up
the cover to remove.
(2)
Insert the lamp clamp lever to open the socket pin
hole. Attach the lamp while holding down the lever.
Put the lamp clamp lever back to its original position.
Designated lamp: PHILIPS7724
(3)
Reattach the cover back to its original position and
tighten the lamphouse cover fixing screw.
4

Attach the lamphouse for dia-illumination.

Tool: Hex driver (2 mm across flats)
(1)
Insert the lamphouse at the rear of the microscope
and tighten the fixing screws.
(2)
Connect the dia-illumination lamphouse cable to the
LAMP connector on the rear of the main body.
Chapter 1
or OSRAM HLX64623
5
Assembly
Insert the hex wrench and
loosen the fixing screw.
Fixing screw
Removing the lamphouse cover
Lamp
Lamphouse
fixing screw
Attaching the lamp
Lamphouse fixing screw
Securing the dia-illumination lamphouse
Lamphouse
cover
Lamp
clamp lever