Nikon Eclipse Ci-E Instruction Manual page 34

Upright motorized microscope
Hide thumbs Also See for Eclipse Ci-E:
Table of Contents

Advertisement

Microscopy operation
14
Bring a portion of the specimen where there is no sample into the optical path.
Rotate the stage knob to move the specimen and bring a portion where there is no sample under the cover glass into
the optical path.
15
Bring the analyzer and the polarizer into the optical path.
Push in the analyzer IN/OUT knob into the analyzer tube
for simple polarization to bring the analyzer into the optical
path.
Set the polarizer unit for simple polarization over the field
lens.
Make sure the orientation mark on the polarizer (roughly)
comes to the front at this point.
Keep the fixing screw of the polarizer unit loosened.
Chap. 1-3
Chapter 1
Slider-type analyzer for simple
polarization
Using the D-SA Analyzer Slider for Simple
Polarization instead of an analyzer tube for simple
polarization can keep the eye point from rising. To
use the D-SA Analyzer Slider for Simple
Polarization, insert it into the analyzer slot of the
nosepiece. Use the analyzer slider as follows:
Push in:
The analyzer goes into the optical
path.
Pull out:
The analyzer is removed from and the
dummy hole goes into the optical path
(empty).
Microscopy Procedures
Pivot
Orientation
mark
Bringing the analyzer into the optical path
D-SA analyzer
slider
Using D-SA analyzer slider
22
Analyzer
IN/OUT knob
0.9
0.8
0.5
0.4
0.3
0.2
0.1
Fixing screw
Installing the polarizer

Advertisement

Table of Contents
loading

Table of Contents