Epi-Fluorescence Microscopy - Nikon Eclipse Ci-E Instruction Manual

Upright motorized microscope
Hide thumbs Also See for Eclipse Ci-E:
Table of Contents

Advertisement

Capture button
Problem
The capture button does
not function.

Epi-fluorescence Microscopy

2.2
Problem
The mercury lamp does
not work.
The mercury lamp burns
out soon after it is turned
on.
Chap. 4
Chapter 4
Cause
A camera trigger cable is not properly
connected.
A destination of a capture trigger signal has
not been set correctly.
Cause
There is no power supplied.
The lamp has burned out.
The mercury lamp is not attached.
The mercury lamp's connector is not
connected to the illuminator.
The lamp type is incorrect.
The lamp is at end of its life.
Troubleshooting
Attach it correctly.
(→Chapter 3, "7 Attaching a Camera".)
Change the setting with DS-L4 or Ni Setup
Tool.
Plug in the power cord.
(→Check your illuminator's manual.)
Replace the lamp with the specified type.
(→Check your illuminator's manual.)
Attach a designated lamp.
(→Check your illuminator's manual.)
Connect it to the illuminator.
(→Check your illuminator's manual.)
Replace the lamp with the specified type.
(→Check your illuminator's manual.)
Contact your nearest Nikon representative for
advice if a new lamp will burn out soon after
replacement.
86
Measure
Measure

Advertisement

Table of Contents
loading

Table of Contents