Nikon Eclipse Ci-E Instruction Manual page 75

Upright motorized microscope
Hide thumbs Also See for Eclipse Ci-E:
Table of Contents

Advertisement

Brightness of fluorescent image
Generation of autofluorescence
Degree of color fading
Barrier filter (BA filter)
Barrier filters allow only fluorescent light emitted by the specimen to pass, blocking excitation light. This allows the
fluorescent image to be viewed without excess illumination (dark background).
There are two types of barrier filters: LP filters that block all light below a certain wavelength but pass all light of longer
wavelengths, and BP filters that pass light of a certain waveband and block all other light. Use the filter type appropriate
for your intended purpose.
LP filter (long-pass filter)
LP filters block all light below a certain wavelength but pass all
light of longer wavelengths. The border wavelength is called the
cut-on wavelength.
(1)
For specimens labeled with a fluorophore in which the
fluorescent waveband and excitation waveband (light that the
specimen absorbs in order to emit fluorescent light) are very
close, selecting a barrier filter with the shortest cut-on
wavelength permitted by the performance requirements will
result in most efficient fluorescent microscopy. If the cut-on
wavelength is long, excitation light and fluorescent light will
be entirely distinct, tending to darken the background of
fluorescent images. However, recent developments in filter
performance have resulted in increased use of filters of short
cut-on wavelengths.
(2)
For multiple-labeled specimens, use an LP filter for
microscopy of fluorescent images of all fluorophores. Note
that a combination involving an ordinary dichroic mirror, an
excitation filter, and an LP-filter-type barrier filter will be
incapable of excited fluorophores that emit long-wavelength
fluorescent light – for example, TRITC in the case of FITC
and TRITC. This will result in very dark TRITC fluorescent
images. For such cases, we recommend using multiband
filters.
BP filter (bandpass filter)
Bandpass filters pass only light of a certain wavelength range,
blocking all other light.
BP filters are used for microscopy of fluorescent images involving
a specific fluorophore in multiple-labeled specimens. (For
example, in a double-labeled specimen of FITC and TRITC, the
BA520-560 filter enables microscopy of just the FITC fluorescent
image.)
However, BP filters will not separate autofluorescence, if any
(because the fluorescent image in the above combination is
green only). LP filters are better suited for making fine separation
of autofluorescence based on slight color differences.
Chapter 2
Individual Operations
EX Filter Bandwidth and Fluorescent Image
Narrow
Dark
Low
Low
63
EX Filter Bandwidth
Waveband
emitted by
FITC
Wavelength
Long-pass filter
(Both FITC and TRITC images are visible.)
BA520-560 (BP type)
Waveband
emitted by
FITC
Wavelength
Band-pass filter
(Only the FITC fluorescent image is visible.)
Wide
Bright
High
High
LP520
Waveband
emitted by
TRITC
Chap. 2
Waveband
emitted by
TRITC

Advertisement

Table of Contents
loading

Table of Contents