Commissioning
Parameter
Assign current output 1
Current span
0/4 mA value
46
10.4.3
Configuring the current output
The Current output submenu guides you systematically through all the parameters that
have to be set for configuring the current output.
Navigation
"Setup" menu → Current output 1
Structure of the submenu
‣
Current output 1
Assign current output 1
Current span
0/4 mA value
20 mA value
Fixed current
Damping output 1
Failure mode
Failure current
Parameter overview with brief description
Prerequisite
–
–
One of the following options is
selected in the Current span
parameter (→ 46):
• 4...20 mA NAMUR
• 4...20 mA US
• 4...20 mA
• 0...20 mA
Description
Select process variable for
current output.
Select current range for
process value output and
upper/lower level for alarm
signal.
Enter 4 mA value.
Proline Prosonic Flow E 100 HART
→ 46
→ 46
→ 46
→ 47
→ 47
→ 47
→ 47
→ 47
Selection / User
Factory setting
entry
• Off
Volume flow
• Volume flow
• Mass flow
• Sound velocity
• Flow velocity
*
• Temperature
*
• Acceptance rate
*
• Signal strength
• Signal to noise
*
ratio
*
• Turbulence
*
• Signal asymmetry
• 4...20 mA NAMUR
Country-specific:
• 4...20 mA US
• 4...20 mA NAMUR
• 4...20 mA
• 4...20 mA US
• 0...20 mA
• Fixed current
Signed floating-point
0 l/h
number
Endress+Hauser