Commissioning
Parameter
Operating mode
Assign frequency output
Minimum frequency value
Maximum frequency value
50
Parameter overview with brief description
Prerequisite
–
The Frequency option is
selected in the Operating
mode parameter (→ 49)
parameter.
In the Operating mode
parameter, the Frequency
option is selected, and one of
the following options is
selected in the Assign
frequency output parameter
(→ 50):
• Volume flow
• Mass flow
• Sound velocity
• Flow velocity
• Temperature
• Acceptance rate
• Signal strength
• Signal to noise ratio
• Turbulence
• Signal asymmetry
In the Operating mode
parameter, the Frequency
option is selected, and one of
the following options is
selected in the Assign
frequency output parameter
(→ 50):
• Volume flow
• Mass flow
• Sound velocity
• Flow velocity
• Temperature
• Acceptance rate
• Signal strength
• Signal to noise ratio
• Turbulence
• Signal asymmetry
Description
Define the output as a pulse,
frequency or switch output.
Select process variable for
frequency output.
Enter minimum frequency.
Enter maximum frequency.
Proline Prosonic Flow E 100 HART
Selection / User
Factory setting
entry
• Pulse
Pulse
• Frequency
• Switch
• Off
Off
• Volume flow
• Mass flow
• Sound velocity
• Flow velocity
• Temperature
• Acceptance rate
• Signal strength
• Signal to noise
ratio
• Turbulence
• Signal asymmetry
0.0 to 10 000.0 Hz
0.0 Hz
0.0 to 10 000.0 Hz
10 000.0 Hz
Endress+Hauser