Output - Endress+Hauser Proline Prosonic Flow E 100 HART Operating Instructions Manual

Ultrasonic time-of-flight flowmeter
Table of Contents

Advertisement

Technical data
Operable flow range
Output signal
92
Flow characteristic values in US units
Nominal
Recommended
diameter
min./max. full scale value
[in]
[mm]
[gal/min]
2
50
0 to 190
2 ½
65
0 to 317
3
80
0 to 444
4
100
0 to 761
6
150
0 to 1 680
Recommended measuring range
"Flow limit" section →  98
Over 200:1
16.4

Output

Current output
Current output
Maximum output values
Load
Resolution
Damping
Assignable measured
variables
Pulse/frequency/switch output
Function
Version
Maximum input values
Voltage drop
Pulse output
Pulse width
Maximum pulse rate
Pulse value
Assignable measured
variables
Frequency output
Output frequency
flow
Full scale value current
output
[gal/min]
1 680
4-20 mA HART (active)
• DC 24 V (no flow)
• 22.5 mA
0 to 700 Ω
0.38 µA
Adjustable: 0.07 to 999 s
• Volume flow
• Mass flow
• Temperature
Can be set to pulse, frequency or switch output
Passive, open collector
• DC 30 V
• 25 mA
For 25 mA: ≤ DC 2 V
Adjustable: 0.05 to 2 000 ms
10 000 Impulse/s
Adjustable
• Volume flow
• Mass flow
Adjustable: 0 to 10 000 Hz
Proline Prosonic Flow E 100 HART
Factory settings
Pulse value
[gal/pulse]
190
0.8
317
1.1
444
1.6
761
2.6
6.6
Endress+Hauser
Low flow cut off
(v ~ 0.1 m/s)
[gal/min]
3.8
6.3
8.9
15.2
33.6