Ensuring The Degree Of Protection; Post-Connection Check - Endress+Hauser Proline Prosonic Flow E 100 HART Operating Instructions Manual

Ultrasonic time-of-flight flowmeter
Table of Contents

Advertisement

Electrical connection
26
Switch output
1
 9
Connection example for switch output (passive)
1
Automation system with switch input (e.g. PLC)
2
Power supply
3
Transmitter: Observe input values
7.4

Ensuring the degree of protection

The measuring device fulfills all the requirements for the IP66/67 degree of protection,
Type 4X enclosure.
To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following
steps after the electrical connection:
1. Check that the housing seals are clean and fitted correctly.
2. Dry, clean or replace the seals if necessary.
3. Tighten all housing screws and screw covers.
4. Firmly tighten the cable glands.
5. To ensure that moisture does not enter the cable entry:
Route the cable so that it loops down before the cable entry ("water trap").
6. Insert dummy plugs into unused cable entries.
7.5

Post-connection check

Are cables or the device undamaged (visual inspection)?
Do the cables used meet the requirements→  21?
Do the cables have adequate strain relief?
Are all the cable glands installed, firmly tightened and leak-tight? Cable run with "water trap"
→  26?
Does the supply voltage match the specifications on the transmitter nameplate →  94?
If supply voltage is present, is the power LED on the electronics module of the transmitter lit green
→  11?
Depending on the device version, is the securing clamp or fixing screw firmly tightened?
Proline Prosonic Flow E 100 HART
2
3
A0028760
A0029278