Pending Diagnostic Events - Endress+Hauser Proline Prosonic Flow E 100 HART Operating Instructions Manual

Ultrasonic time-of-flight flowmeter
Table of Contents

Advertisement

Diagnostics and troubleshooting
80
Diagnostic
Short text
number
Diagnostic of process
803
Current loop
832
Electronic temperature
too high
833
Electronic temperature
too low
834
Process temperature too
high
835
Process temperature too
low
840
Sensor range
842
Process limit
881
Sensor signal path 1 to n 1. Check process conditions
882
Input signal
1)
Diagnostic behavior can be changed.
12.7

Pending diagnostic events

The Diagnostics menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via Web browser →  75
• Via "FieldCare" operating tool →  76
• Via "DeviceCare" operating tool →  76
Other pending diagnostic events can be displayed in the Diagnostic list submenu
→  81
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics
Previous diagnostics
Operating time from restart
Operating time
Proline Prosonic Flow E 100 HART
Remedy instructions
1. Check wiring
2. Change I/O module
Reduce ambient temperature
Increase ambient temperature
Reduce process temperature
Increase process temperature
Check flow velocity
Low flow cut off active!
1. Check low flow cut off
configuration
2. Replace sensor electronic module
(ISEM)
3. Replace device
1. Check input configuration
2. Check external device or process
conditions
Status
Diagnostic
signal
behavior
[from the
[from the
factory]
factory]
F
Alarm
1)
S
Warning
1)
S
Warning
1)
S
Warning
1)
S
Warning
1)
S
Warning
S
Warning
1)
M
Warning
F
Alarm
→  81