Overview
IMPORTANT
the USB configuration key is not connected, the following error message appears: CytExpert cannot find
the license. Please check whether the correct USB configuration key has been plugged in.
This chapter describes the instrument startup procedure.
Workflow:
Pre-startup
inspection
This chapter contains information on:
Pre-Startup Inspection
•
•
Turning On the Instrument
Logging Into the Software
•
•
System Startup Program
•
Selecting Experiments from the Start Page
Pre-Startup Inspection
Before using the CytoFLEX SRT instrument, perform the following system checks.
System Connections Inspection
1
Verify that the power cable of the computer is securely connected to the power source.
2
Check that the monitor, mouse, keyboard, and the Sorter USB cable are properly connected to
the computer. Refer to
C37808AC
Verify that the correct USB configuration key is securely connected to a computer USB port. If
Turn on power
Ô
Figure
1.34.
Daily Startup
Open software
Ô
Ô
CHAPTER 3
System Startup
Program
3-1