Open the sort chamber sliding door and the sort protection door.
3
Clean the bottom of the flow cell, the sliding rails for the nozzle lift, and the V-plate surface. For
instructions, refer to
Procedures.
IMPORTANT
4
Verify that the nozzle is clean and dry. If necessary, clean the nozzle. For instructions, refer to
Cleaning the Nozzle
C37808AC
Daily Decontamination During Shutdown
A wet or clogged nozzle might cause the fluid stream off center or even no stream.
in
CHAPTER 10, Cleaning
in
CHAPTER 10, Cleaning
Procedures.
3
Daily Startup
System Startup Program
3-9