a. Windows 10 (2019 LTSC) is the only operating system that has been validated to work with the CytoFLEX SRT workstation.
b. If you need connect an external device or USB to the computer, Beckman Coulter recommends using the wireless keyboard
and/or wireless mouse for the computer. Otherwise, you need at least 6 interfaces to ensure one can be used for the
external device or USB when using the wired keyboard and the wired mouse.

Performance Characteristics

Sensitivity
Fluorescence resolution
Side scatter resolution
C37808AC
Electronics
7-decade data display
Pulse area and height for every channel, width for two selectable channels
Data Management
CytExpert SRT software
English and Chinese
FCS 3.0
Operating system
Windows
Processor
Intel
or Intel
Monitor
2560 x1440 32 inch
Memory
8 GB RAM or higher
Storage
1 TB or higher
Ethernet
integrated 100 MB, Dual Ethernet ports
USB interface
≥ 5
Full matrix compensation, manual and automatic.
Novel Compensation Library for storage of spillover values of dyes to easily
determine the correct compensation matrix with new gain settings.
Performance
MESF
FITC: < 30 molecules of equivalent soluble fluorochrome
(MESF-FITC)
PE: < 10 molecules of equivalent soluble fluorochrome
(MESF-PE)
APC: < 25 molecules of equivalent soluble fluorochrome
(MESF-APC)
rCV < 3%
The CytoFLEX SRT cell sorter is capable of achieving <3% rCV. Using
CytoFLEX Daily QC Fluorospheres or CytoFLEX Ready to Use Daily QC
Fluorospheres for daily QC, the pass criteria is ≤ 5%.
< 300 nm
Performance Characteristics
®
10 Enterprise LTSC 2019 x64-bit
®
Core™ i7-7700 (8MB Cache, 3.6 GHz;
®
Core™ i7-8700T 12MB Cache, 2.4 GHz)
b
interfaces, at least one USB 3.0 interface
1
System Overview
a
1-43