Beckman Coulter CytoFLEX SRT Instructions For Use Manual page 171

Cell sorter
Table of Contents

Advertisement

Instrument Quality Control and Standardization
Quality Control
2
Select a QC run from the QC Process list on the left and a QC report appears on the right.
NOTE
The results column indicates a passing QC result with a
QC results must meet the following criteria to pass:
4-14
DD1 must be within the range of 720 - 880 μs.
DD laser must be within the range of 90 - 110 mW.
The delay differences must be 5 μs.
The gain differences must be 20% from the target gain.
The median fluorescence intensity (MFI) differences must be 5% from the target MFI.
The rCV must be within the range.
and a failed QC result with