Running Shutdown Program; System Shutdown - Beckman Coulter CytoFLEX SRT Instructions For Use Manual

Cell sorter
Table of Contents

Advertisement

Shutting Down the System

Running Shutdown Program

2
Run Daily Clean to clean the sample line. Refer to
Procedures.
3
Select
Standby
Running Shutdown Program
The System Shutdown Program is to rinse the sample line and perfuse the flow cell with the
CytoFLEX SRT Shutdown fluid to prevent saline accumulation.
It is required to run the System Shutdown Program every time you shut down the instrument. Refer
to
System
Shutdown. However, if the instrument will not be used for more than 7 days, it is
recommended to perform the Long Term Shutdown program. Refer to

System Shutdown

1
Select
Standby
2
Select
System Shutdown
8-2
to Turn off Sheath.
to turn off the sheath.
from the Cytometer menu.
Daily Clean Program
in
CHAPTER 10, Cleaning
Long Term
Shutdown.
C37808AC