The report area on the right displays detailed experiment results, including laser power, delay,
testing conditions, and signal results. The same
result. For items that fail, values falling outside the prescribed range are displayed in red font.
In the Comment area, an explanation appears for each failed item.
3
If QC fails, follow the procedure below:
a. Verify whether the beads used were within their shelf life and stored in accordance with
the appropriate instruction manual.
b. Verify whether the allocated sample tube was prepared as required and correctly
positioned.
c. Verify whether the detector configuration selected for QC matches with the current
detector configuration. Refer to
Configuration
d. Run
Procedures, and retest.