Beckman Coulter CytoFLEX SRT Instructions For Use Manual page 47

Cell sorter
Table of Contents

Advertisement

1.
Wash station.
backflush.
2.
Safety interlock.
is opened.
3.
LED indicator.
• Green: Sorter is in either idle, or ready, or standby state.
• Red: A critical error has occurred
• Yellow: The system requires the operator's attention.
• Flashing green: Certain program is in process, for example, loading sample, QC, etc.
4.
Sample chamber.
to form a closed and pressurized chamber.
NOTE
Otherwise, the system prompts an error.
5.
Sample station door.
movement.
6.
Sample tube holder.
7.
Sample probe
8.
Tube light
CAUTION
Risk of saline buildup in the fluidics system. Do not use the Urgent Stop button for
routine shutdown. Urgent stop skips the important cleaning steps and may cause
high carryover.
9.
Urgent Stop
NOTE
• Sample loading stopped
• Sheath flow off
• Waste catcher extended
• Deflection plate voltage off
• Lasers off.
NOTE
Refer to
Program prior to perform the acquisition or sorting.
10.
Backflush tubing.
liquid to the waste container.
C37808AC
Automatically cleans the sample probe when the instrument performs a
Stops the sample chamber movement when the sample station door
Indicates the working status of the Sorter.
During the sampling process, the sample chamber cover moves down
Do not open the sample station door when the sample chamber cover is moving.
Engages or disengages the safety interlock of the sample chamber
Supports 12 x 75 mm tubes for sample acquisition.
. Draws and transfers sample into the flow cell.
. Illuminates the sample tube.
. Puts the system into the idle state in urgent cases.
Long-pressing this button puts the instrument in the following states:
After using the Urgent Stop button, follow the Daily Startup procedures to resume use.
CHAPTER 3, Daily
Startup. Beckman Coulter recommends running Shutdown
Provides fluid to clean the sample probe, and conveys the waste
1
System Overview
Fluidics System
1-15