Shutting Down the System
Running Shutdown Program
IMPORTANT
leaving the instrument. The entire process takes about 30 minutes.
IMPORTANT
Empty the sheath tank and refill the sheath tank with IsoFlow sheath.
Switch the aseptic cleaning filter back to the sheath fluid filter
Restart the instrument
Run the System Startup Program.
Required materials
• 1.5 L 70% ethanol in deionized water
• An aseptic clean solution filter
• A new waste air filter
1
Run System Shutdown program to put the system in idle state. Refer to
IMPORTANT
Otherwise, you could encounter System Startup failure, Aseptic Cleaning failure, Long-term Shutdown
failure, or sheath flow indication error.
2
Verify that the sheath tank you used are matched with the sheath tank type you selected. Select
Sheath Tank Scale Reset from the Advanced menu to display the current sheath tank type.
3
Select
Long Term Shutdown
8-8
Once you begin, ensure that you complete the entire Long Term Shutdown procedure before
If the Long Term Shutdown procedure was terminated by accident, perform the following:
Ensure that the sheath tank you used matches with the sheath tank type you selected.
in the Cytometer menu.
System
Shutdown.