Sorting
Sorting

Viewing Sorting Statistics

CAUTION
Risk of sample loss and biohazardous contamination. Remove the sample
collection tube when the tube is nearly full. The Tube Volume value on the
software might not be accurate due to the various sample conditions. Use suitable
laboratory attire when removing the sample.
Table 5.2 Sort Statistics
Screen Element
Sort Count
Sort Rate
Abort Count
Abort Rate
% Total
Efficiency
Total Volume
5-102
Function
Total positive events that have been sorted for the stream.
Total sorted events per second for the stream.
Total positive events that have been aborted for the stream.
Total aborted events per second for the stream.
The percent of positive sorted events relative to the Total Events for the
acquisition.
The number of positive events sorted, divided by the total events that could have
been sorted for the stream. sorted/(sorted + aborted)
The percent of the collection volume relative to the tube size. (number of charged
drops x droplet volume + initial volume)/tube volume
• Grey: indicates the initial volume
• Green: indicates that the collection volume is normal
• Red: indicates that the collection volume reaches the limit. You need pause
sorting and replace a new tube.
NOTE
This Total Volume is based on theoretical estimation and may be
inaccurate. Check the status of sample collection tube on a regular basis and
replace the tube if it is nearly full.