Beckman Coulter CytoFLEX SRT Instructions For Use Manual page 388

Cell sorter
Table of Contents

Advertisement

Table 9.1 Troubleshooting-[Error Codes] (Continued)
Problem
[Error code 060063] Failed to
determine the Detector Delay
for L2 stream.
[Error code 060064] Failed to
determine the Detector Delay
for L1 stream.
[Error code 060065] Failed to
determine the Detector Delay
for R1 stream.
[Error code 060066] Failed to
determine the Detector Delay
for R2 stream.
[Error code 060067] Failed to
determine the Detector Delay
for Gap detector.
[Error code 060068] Failed to
determine the Detector Delay
for Central stream.
C37808AC
Probable Cause
• The fluid stream is unstable.
• The photodiodes are defective.
• Airflow interference.
• Electric leakage of high-voltage
component.
• Leakage of the fluidics system.
• The narrow slot mask of the PD is
partly or completely blocked.
• The core stream or the side stream
module is not aligned.
Corrective Action
1. Select Standby to turn off the sheath.
2. Clean the side stream illumination and the deflection
plates. Refer to
Cleaning the Side Stream
Illumination Source and the Deflection Plates
CHAPTER 10, Cleaning
Procedures.
3. Wipe the side stream detection window. Refer to
Figure
1.23.
4. Verify that the nozzle is clean and O-ring is present.
• Clean the nozzle if needed. Refer to
Nozzle
in
CHAPTER 10, Cleaning
• If necessary, replace the O-ring. Refer to
Replacing the O-Ring
in
Replacement/Adjustment
• If the nozzle is damaged, replace the nozzle.
Refer to
Replacing the Nozzle
Replacement/Adjustment
5. Reinstall the nozzle module. Refer to
Installing the Nozzle Module
Replacement/Adjustment
Procedures.
6. Run Flow Cell Debubble. Refer to
Flow Cell De-bubble
in
CHAPTER 11, Replacement/
Adjustment
Procedures.
7. Run Sheath Filter debubble. Refer to
Sheath Filter De-bubble
in
Replacement/Adjustment
Procedures.
8. Run Daily Clean. Refer to
Daily Clean Program
CHAPTER 10, Cleaning
Procedures.
9. Run Flow cell Clean. Refer to
Program
in
CHAPTER 10, Cleaning
10. Verify that the Sorter is not exposed directly to a
ventilation outlet of an air condition.
11. Rerun Sort Calibration. Refer to
(Auto Drop Delay)
in
CHAPTER 5,
12. Restart the system. Refer to
Startup.
13. If the problem persists,
contact
9
Troubleshooting
Troubleshooting Table
in
Cleaning the
Procedures.
CHAPTER 11,
Procedures.
in
CHAPTER 11,
Procedures.
Removing/
in
CHAPTER 11,
Performing the
Performing the
CHAPTER 11,
in
Flow Cell Clean
Procedures.
Sort Calibration
Sorting.
CHAPTER 3, Daily
us.
9-17