Sorting
Load Sample and Record Data

Verifying, Selecting, Editing, and Creating Detector Configuration

CAUTION
Risk of erroneous results. The system will read the selected Detector
Configuration even if the optical filters do not match the selected Detector
Configuration. You must verify the installed optical filters match the selected
Detector Configuration.
1
Select
Detector Configuration
configuration is selected. To change the configuration:
a. Select the desired configuration.
b. Select
A green check mark appears in front of the selected configuration.
NOTE
A configuration is locked when
two reasons:
Locked configurations can be deleted, but not edited.
2
Select
OK
3
Proceed to Step
need to create a new Detector Configuration, or skip to Step
configuration.
5-60
in the Cytometer menu to verify the correct detector
.
Set as Current
QC was run using the configuration.
The compensation library contains data for the configuration.
to close the Detector Configuration screen.
4
if you need to edit the Detector Configuration settings, or skip to Step
appears to the left of a configuration. A configuration locks for
12
if you need to delete a Detector
5
if you
C37808AC