Performance Specifications; Conformance To Specification (±3S [Sigma]); Reference Accuracy - Emerson Rosemount 3051S Series Reference Manual

Pressure transmitter with foundation fieldbus protocol
Hide thumbs Also See for Rosemount 3051S Series:
Table of Contents

Advertisement

Specifications and Reference Data
October 2016
Appendix A
Performance specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 99
Functional specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 107
Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 112
Dimensional drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 115
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 123
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 124
Exploded view diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 153
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 154
A.1 Performance specifications
For zero-based spans, reference conditions, silicone oil fill,
glass-filled PTFE O-rings, SST materials, coplanar flange
(Rosemount
3051SMV, 3051S_C) or
3051S_T) process connections, digital trim values set to equal
range points.
Transmitter with Coplanar Sensor Module (Single Variable)
Differential Pressure (Rosemount 3051S_CD), Gage Pressure (Rosemount 3051S_CG)
Ultra
±0.025% of span;
Ranges
For spans less than 10:1,
2 - 4
±(0.005 + 0.0035 [URL/Span])% of span
±0.05% of span;
Range 5
For spans less than 10:1,
±(0.005 + 0.0045 [URL/Span])% of span
±0.09% of span;
Range 1
For spans less than 15:1,
±(0.015 + 0.005 [URL/Span])% of span
±0.09% of span;
Range 0
For spans less than 2:1, ±0.045% of URL
Absolute Pressure (Rosemount 3051S_CA)
Ultra
±0.025% of span;
Ranges
For spans less than 10:1,
1 - 4
±(0.004 [URL/Span])% of span
±0.075% of span;
Range 0
For spans less than 5:1,
±(0.025 + 0.01 [URL/Span])% of span
1.
Ultra for Flow is only available for Rosemount 3051S_CD ranges 2-3 and Rosemount 3051SMV DP ranges 2-3. For calibrated spans from 1:1 to 2:1 of URL, add ±0.005%
of span analog output error.
Specifications and Reference Data
Specifications and Reference Data
1
/
in.- 14 NPT (Rosemount
2
Classic
±0.035% of span;
For spans less than 10:1,
±(0.015 + 0.005 [URL/Span])% of span
±0.065% of span;
For spans less than 10:1,
±(0.015 + 0.005 [URL/Span])% of span
±0.10% of span;
For spans less than 15:1,
±(0.025 + 0.005 [URL/Span])% of span
±0.10% of span;
For spans less than 2:1, ±0.05% of URL
Classic
±0.035% of span;
For spans less than 10:1,
±(0.0065 [URL/Span])% of span
±0.075% of span;
For spans less than 5:1,
±(0.025 + 0.01 [URL/Span])% of span
A.1.1 Conformance to specification (±3 [Sigma])
Technology leadership, advanced manufacturing techniques, and
statistical process control ensure measurement specification
conformance to ±3or better.
A.1.2 Reference accuracy
Stated reference accuracy equations include terminal based
linearity, hysteresis, and repeatability.
For F
Fieldbus and wireless devices, use calibrated
OUNDATION
range in place of span.
Ultra for Flow
±0.04% of reading up to 8:1 DP
turndown from URL;
±(0.04 + 0.0023 [URL/Reading])% of
reading to 200:1 DP turndown from URL
Not available
Not available
Not available