Refer to the Rosemount 3051S HART Reference Manual for more instruction. This document is also available electronically on Emerson.com/Rosemount. WARNING Explosions could result in death or serious injury. • Do not remove the transmitter cover in explosive atmospheres when the circuit is live.
February 2019 Quick Start Guide Mount the transmitter Liquid flow applications Procedure 1. Place taps to the side of the line. 2. Mount beside or below the taps. 3. Mount the transmitter so that the drain/vent valves are oriented upward. A.
Page 4
™ Emerson provided mounting brackets. Use only bolts provided with the transmitter or sold as Emerson spare parts. 1.4.1 Mounting brackets Figure 1-1: Mounting Bracket – Coplanar Flange Panel mount...
Page 5
February 2019 Quick Start Guide Figure 1-2: Mounting Brackets – Traditional Flange Panel mount Pipe mount Figure 1-3: Mounting Brackets – In-line Panel mount Pipe mount Quick Start Guide...
Page 6
Only use bolts ™ supplied with the transmitter or sold by Emerson as spare parts. Figure 1-4 illustrates common transmitter assemblies with the bolt length required for proper transmitter assembly.
Page 7
Table 1-1 . If bolt material is not shown in Table 1-1, contact the local Emerson representative for more information. Use the following bolt installation procedure: Procedure 1. Carbon steel bolts do not require lubrication and the stainless steel bolts are coated with a lubricant to ease installation.
Page 8
Quick Start Guide February 2019 Figure 1-5: Proper Bolt Installation A. Bolt B. Sensor module O-rings with flange adapters WARNING Failure to install proper flange adapter O-rings may cause process leaks, which can result in death or serious injury. Only use the O-ring that is designed for its specific flange adapter.
Page 9
February 2019 Quick Start Guide Keep the vent path free of any obstruction, including but not limited to paint, dust, and lubrication by mounting the transmitter so that any contaminants can drain away. Figure 1-6: In-line Gage Transmitter A. Low side pressure port (under neck label) Quick Start Guide...
Figure 2-1: Commissioning Tag for FOUNDATION Fieldbus revision ITK5 (default option) COMMISSIONING TAG Device ID: 0011513051010001440-121698091725 PD Tag: PT- 101 Revision: 23 Support files available at Emerson.com/Rosemount Tear Here Revision: 23 Support files available at Emerson.com/Rosemount Device Serial Number: XXXXXXXXXX Device ID: 0011513051010001440-121698091725...
Page 11
Figure 2-2: Commissioning Tag FOUNDATION Fieldbus revision ITK6 (option code IT6) COMMISSIONING TAG Device ID: 001151AA00010001440-121698091725 PD Tag: PT- 101 Revision: 24 Support files available at Emerson.com/Rosemount Tear Here Revision: 24 Support files available at Emerson.com/Rosemount Device Serial Number: XXXXXXXXXX Device ID: 001151AA00010001440-121698091725...
Quick Start Guide February 2019 Consider housing rotation To improve field access to wiring or to better view the optional LCD display: Procedure 1. Loosen the housing rotation set screw. 2. Turn the housing up to 180° left or right of its original (as shipped) position.
February 2019 Quick Start Guide Connect wiring and power up Procedure 1. Remove and discard orange conduit plugs. 2. Remove the housing cover labeled “Field Terminals.” Note Do not connect the power across the test terminals. Power could damage the test diode in the test connection. Twisted pairs yield best results.
Page 14
Quick Start Guide February 2019 Figure 4-1: Transmitter Wiring (Plantweb Housing) A. RL ≥ 250 Ω B. Power supply Figure 4-2: Transmitter Wiring (Junction Box Housing) A. RL ≥ 250 Ω B. Power supply Note Installation of the transient protection terminal block does not provide transient protection unless the Rosemount 3051S case is properly grounded.
Page 15
February 2019 Quick Start Guide • Continuously connect to the termination point • Be connected to a good earth ground at the power supply end Figure 4-3: Wiring A. Insulate shield B. Minimize distance C. Connect shield back to the power supply ground D.
Page 16
Quick Start Guide February 2019 Cable type Recommend Madison AWM Style 2549 cable. Other comparable cable may be used as long as it has independent dual twisted shielded pair wires with an outer shield. The power wires must be 22 AWG minimum and the CAN communication wires must be 24 AWG minimum.
Page 17
February 2019 Quick Start Guide Figure 4-4: Remote Mount Display Wiring Diagram (white) 24 AWG (blue) 24 AWG (black) 22 AWG (red) 22 AWG A. Remote mount display B. Junction box housing C. 4–20 mA Note Wire colors provided are per Madison AWM Style 2549 cable. Wire color may vary depending on cable selected.
Page 18
Quick Start Guide February 2019 Figure 4-5: Rosemount 3051S Quick Connect Exploded View A. Straight field wireable connector (part number 03151-9063-0001), supplied by customer B. Right angle field wireable connector (part number 03151-9063-0002), supplied by customer C. Quick Connect housing D.
Page 19
February 2019 Quick Start Guide Figure 4-6: Quick Connect Housing Pin-Out “+” “ ” A. Ground B. No connection Note For other wiring details, refer to pin-out drawing and the cordset manufacturer’s installation instructions. Conduit electrical connector wiring (option GE or GM) For Rosemount 3051S with conduit electrical connectors GE or GM, refer to the cordset manufacturer’s installation instructions for wiring details.
Page 20
Quick Start Guide February 2019 Figure 4-7: Load Limitation 1387 1000 Operating Region 10.5 42.4 Voltage (Vdc) Maximum loop resistance = 43.5 x (power supply voltage – 10.5) The Field Communicator requires a minimum loop resistance of 250Ω for communication. Rosemount 3051S...
February 2019 Quick Start Guide Verify configuration Use the following block examples to do basic configuration to the transmitter. ™ For more advanced configurations see the Rosemount 3051S F OUNDATION Fieldbus Reference Manual. Note ™ DeltaV users should use DeltaV Explorer for the resource and transducer blocks and Control Studio for the function blocks.
Page 22
Quick Start Guide February 2019 Parameters Enter data Out_Scale Set values outside operating range. DP Flow example Parameters Enter data Channel L_Type Square root XD_Scale 0–100 inH O @ 68 °F Out_Scale 0–20 GPM DP Level example Parameters Enter data Channel L_Type Indirect...
February 2019 Quick Start Guide Trim the transmitter Transmitters are shipped fully calibrated per request or by the factory default of full scale (lower range value = zero, upper range value = upper range limit). Zero trim A zero trim is a single-point adjustment used for compensating mounting position and line pressure effects.
A copy of the EU Declaration of Conformity can be found at the end of the Quick Start Guide. The most recent revision of the EU Declaration of Conformity can be found at Emerson.com/Rosemount. Ordinary Location Certification As standard, the transmitter has been examined and tested to determine that...
Page 25
February 2019 Quick Start Guide Special Condition for Safe Use: 1. The Model 3051S/3051S-ERS Pressure Transmitter contains aluminum and is considered to constitute a potential risk of ignition by impact or friction. Care must be taken into account during installation and use to prevent impact and friction.
Page 27
February 2019 Quick Start Guide environmental conditions to which the diaphragm will be subjected. The manufacturer's instructions for installation and maintenance shall be followed in detail to assure safety during its expected lifetime. 2. Flameproof joints are not intended for repair. 3.
Page 28
Quick Start Guide February 2019 2. The terminal pins of the Model 3051S SuperModule must be provided with a degree of protection of at least IP20 in accordance with IEC/EN 60529. 3. The Model 3051S enclosure may be made of aluminum alloy and given a protective polyurethane paint finish;...
Page 29
February 2019 Quick Start Guide Special Conditions for Safe Use (X): 1. Cable entries must be used which maintain the ingress protection of the enclosure to at least IP66. 2. Unused cable entries must be filled with suitable blanking plugs which maintain the ingress protection of the enclosure to at least IP66.
Page 30
Quick Start Guide February 2019 Special Conditions for Safe Use (X): 1. This device contains a thin wall diaphragm less than 1 mm thickness that forms a boundary between EPL Ga (process connection) and EPL Gb (all other parts of the equipment). The model code and datasheet are to be consulted for details of the diaphragm material.
Page 31
February 2019 Quick Start Guide Table 7-5: Input Parameters SuperModule 30 V 300 mA 1.0 W 30 nF 3051S...A; 3051SF…A; 30 V 300 mA 1.0 W 12 nF 3051SAL…C 3051S…F; 3051SF…F 30 V 300 mA 1.3 W 3051S …A…M7, M8, or 30 V 300 mA 1.0 W...
Page 32
Quick Start Guide February 2019 Table 7-6: Input Parameters (continued) 3051S...A; 3051SF…A; 30 V 300 mA 1.0 W 12 nF 3051SAL…C 3051S…F; 3051SF…F 30 V 300 mA 1.3 W 3051S …A…M7, M8, or 30 V 300 mA 1.0 W 12 nF 60 µH M9;...
Page 33
February 2019 Quick Start Guide Table 7-7: Input Parameters (continued) Inductance L Special Conditions for Safe Use (X): 1. The Model 3051S Transmitters fitted with transient protection are not capable of withstanding the 500 V test as defined in Clause 6.3.13 of EN 60079-11:2012.
Page 34
Quick Start Guide February 2019 7.4.6 N7 IECEx Type n Certificate IECEx BAS 04.0018X Standards IEC 60079-0: 2011, IEC 60079-15: 2010 Markings Ex nA IIC T5 Gc,(–40 °C ≤ T ≤ +85 °C) Special Condition for Safe Use (X): 1. The equipment is not capable of withstanding the 500 V insulation test required by clause 6.5 of EN 60079-15:2010.
Page 35
February 2019 Quick Start Guide Markings Ex ia IIC T4 Ga (–60 °C ≤ T ≤ +70 °C), IP66 Special Conditions for Safe Use (X): 1. The surface resistivity of the antenna is greater than 1 GΩ. To avoid electrostatic charge buildup, it must not be rubbed or cleaned with solvents or a dry cloth.
Page 36
Quick Start Guide February 2019 3051S-ERS: GJY15.1406X Standards 3051S: GB3836.1-2010, GB3836.2-2010, GB3836.20-2010, GB12476.1-2013, GB12476.5-2013 3051SFx: GB3836.1-2010, GB3836.2-2010, GB3836.20-2010, GB12476.1-2013, GB 12476.5-2013 3051S-ERS: GB3836.1-2010, GB3836.2-2010, GB3836.20-2010 Markings 3051S: Ex d IIC T6…T4; Ex tD A20 T105 °C T 95 °C; IP66 3051SFx: Ex d IIC T4~T6 Ga/Gb;...
Page 37
February 2019 Quick Start Guide 7.7.2 IM Technical Regulation Customs Union (EAC) Intrinsic Safety Certificate RU C-US.AA87.B.00378 Markings 0Ex ia IIC T4 Ga X 7.7.3 IN Technical Regulation Customs Union (EAC) Intrinsic Safety Certificate: RU C-US.AA87.B.00378 0Ex ia IIC T4 Ga X Markings: Japan 7.8.1 E4 Japan Flameproof...
Page 38
Quick Start Guide February 2019 Republic of Korea 7.9.1 EP Republic of Korea Flameproof Certificate 12-KB4BO-0180X [Mfg USA], 11-KB4BO-0068X [Mfg Singapore] Markings Ex d IIC T6...T4 7.9.2 IP Republic of Korea Intrinsic Safety Certificate 12-KB4BO-0202X [HART - Mfg USA],12-KB4BO-0204X [Fieldbus - Mfg USA], 12-KB4BO-0203X [HART - Mfg Singapore], 13- KB4BO-0296X [Fieldbus - Mfg Singapore] Markings...
Page 39
February 2019 Quick Start Guide 7.11.2 SBV Bureau Veritas (BV) Type Approval Certificate 31910 BV Requirements Bureau Veritas Rules for the Classification of Steel Ships Application Class Notations: AUT-UMS, AUT-CCS, AUT-PORT and AUT- IMS. 7.11.3 SDN Det Norske Veritas (DNV) Type Approval Certificate TAA00000K9 Intended Use Det Norske Veritas' Rules for Classification of Ships, High...
Quick Start Guide February 2019 Rosemount 3051S Declaration of Conformity EU Declaration of Conformity No: RMD 1044 Rev. AD Rosemount Inc. 8200 Market Boulevard Chanhassen, MN 55317-9685 declare under our sole responsibility that the product, Rosemount 3051S Series Pressure Transmitters Rosmeount 3051SFx Series Flowmeter Transmitters Rosemount 300S Housings manufactured by,...
Page 41
February 2019 Quick Start Guide EU Declaration of Conformity No: RMD 1044 Rev. AD EMC Directive (2014/30/EU) Harmonized Standards: EN 61326-1:2013, EN 61326-2-3:2013 PED Directive (2014/68/EU) Rosemount 3051S Series Pressure Transmitters Rosemount 3051S_CA4; 3051S_CD2, 3, 4, 5 (also with P0 & P9 option) Pressure Transmitters QS Certificate of Assessment –...
Page 42
Quick Start Guide February 2019 EU Declaration of Conformity No: RMD 1044 Rev. AD ATEX Directive (2014/34/EU) BAS01ATEX1303X – Intrinsic Safety Certificate Equipment Group II, Category 1 G Ex ia IIC T4 Ga Harmonized Standards Used: EN 60079-0:2012+A11:2013, EN 60079-11:2012 BAS01ATEX3304X –...
Page 43
February 2019 Quick Start Guide EU Declaration of Conformity No: RMD 1044 Rev. AD PED Notified Body DNV GL Business Assurance Italia S.r.l. [Notified Body Number: 0496] Via Energy Park, 14, N-20871 Vimercate (MB), Italy Note – equipment manufactured prior to 20 October 2018 may be marked with the previous PED Notified Body number;...
Quick Start Guide February 2019 China RoHS Rosemount 3051S China RoHS List of Rosemount 3051S Parts with China RoHS Concentration above MCVs / Hazardous Substances Hexavalent Polybrominated Polybrominated Part Name Lead Mercury Cadmium Chromium biphenyls diphenyl ethers (Pb) (Hg) (Cd) (Cr +6) (PBB) (PBDE)
Page 45
February 2019 Quick Start Guide Quick Start Guide...
Page 46
Quick Start Guide February 2019 Rosemount 3051S...
Page 47
February 2019 Quick Start Guide Quick Start Guide...
Page 48
The Emerson logo is a Facebook.com/Rosemount trademark and service mark of Emerson Electric Youtube.com/user/ Co. Rosemount is mark of one of the Emerson RosemountMeasurement family of companies. All other marks are the Google.com/+RosemountMeasurement property of their respective owners.