Reference Manual 00809-0100-4801, Rev EA August 2010 Rosemount 3051S Series Scalable Pressure, Flow, and Level Solution ® with HART Protocol www.rosemount.com...
Page 3
North American Response Center Equipment service needs. 1-800-654-7768 (24 hours—includes Canada) Outside of these areas, contact your local Emerson Process Management representative. The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The following 3051S pressure transmitters and the Rosemount 300S Housing MODELS COVERED Kit are covered in this manual. ™ Rosemount 3051S Coplanar Pressure Transmitter Measurement Type Performance Class Differential Gage...
SERVICE SUPPORT nearest Emerson Process Management representative. Within the United States, call the Emerson Process Management Instrument and Valves Response Center using the 1-800-654-RSMT (7768) toll-free number. This center, available 24 hours a day, will assist you with any needed information or materials.
Page 14
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Electrical shock can result in death or serious injury. • Avoid contact with the leads and terminals. Process leaks could result in death or serious injury. • Install and tighten all four flange bolts before applying pressure.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Side Mounting When the transmitter is mounted on its side, position the Coplanar flange to ensure proper venting or draining. Mount the flange as shown in Figure 2-3 on page 2-11, keeping drain/vent connections on the bottom for gas service and on the top for liquid service.
Page 18
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 2-1. HART Installation Flowchart START HERE Bench Calibration? Field Install Configure Security and Alarm Configure Verify (page 2-13) (Section 3 ) Mount Transmitter Review (page 2-6) Set Units Transmitter...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 For dimensional drawing information refer to Appendix A: Specifications and INSTALLATION Reference Data on page A-16. PROCEDURES Process Flange Orientation Mount the process flanges with sufficient clearance for process connections.
Carbon steel bolts do not require lubrication. No additional lubricant should be applied when installing either type of bolt. Bolts supplied by Emerson Process Management are identified by their head markings: Carbon Steel (CS) Head Markings...
Page 21
Rosemount 3051S Series August 2010 Bolt Installation Only use bolts supplied with the Rosemount 3051S or sold by Emerson Process Management as spare parts. When installing the transmitter to one of the optional mounting brackets, torque the bolts to 125 in-lb.
Page 22
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 GAGE/ABSOLUTE TRANSMITTER DIFFERENTIAL TRANSMITTER Drain/Vent Plug Drain/Vent Drain/Vent 1.75 (44) x 4 1.50 (38) x 2 1.50 (38) x 4 1.75 (44) x 4 NOTE Dimensions are in inches (millimeters).
Page 23
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Impulse Piping The piping between the process and the transmitter must accurately transfer the pressure to obtain accurate measurements. There are five possible sources of error: pressure transfer, leaks, friction loss (particularly if purging is used), trapped gas in a liquid line, liquid in a gas line, density variations between the legs, and plugged impulse piping.
Page 24
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Suite includes: Statistical Process Monitoring (SPM), Power Advisory, Status Log, Variable Log, Advanced Process Alerts, Service Alerts, and Time Stamp capability. The Advanced HART Diagnostics Electronics can be ordered using option code DA2 in the transmitter model number or as a spare part (p/n 03151-9071-0001) to retrofit existing 3051S transmitters in the field.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 2-3. Coplanar Installation Examples LIQUID SERVICE GAS SERVICE STEAM SERVICE FLOW Figure 2-4. In-Line Installation Examples LIQUID GAS SERVICE STEAM SERVICE SERVICE 3051S transmitter flange process connection size is –18 NPT.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 2-5. O-Rings. Failure to install proper flange adapter O-rings may cause process leaks, which can result in death or serious injury. The two flange adapters are distinguished by unique O-ring grooves.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 1. Loosen the housing rotation set screw. 2. Turn the housing up to 180° to the left or right of its original (as shipped) position. NOTE Do not rotate the housing more than 180 degrees without first performing a disassembly procedure (see “Remove Interface Assembly”...
Page 28
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 2-7. Switch and jumper configuration (option D1) PlantWeb Housing Junction Box Housing Switches Jumpers Alarm Alarm Security Security Slide the security and alarm Pull the pins out and rotate 90° into...
Page 29
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Field Communicator 1, 4, 4, 1 Fast Keys 2, 2, 7 Device Dashboard Fast Keys 1. Enter the fast key sequence “Local Keys Control” to bring up the “Field device info” screen.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Use twisted pairs to yield best results. To ensure proper communication, use Connect Wiring and 24 AWG to 14 AWG wire, and do not exceed 5000 feet Power Up (1 500 meters).
Page 31
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Signal Wiring Grounding Do not run signal wiring in conduit or open trays with power wiring, or near heavy electrical equipment. Grounding terminations are provided on the sensor module and inside the Terminal Compartment. These grounds are used when transient protect terminal blocks are installed or to fulfill local regulations.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Electrical Considerations Proper electrical installation is necessary to prevent errors due to improper grounding and electrical noise. For Junction Box housing, shielded signal wiring should be used in high EMI/RFI environments.
Page 33
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 2-10. Remote Mount Display wiring diagram Junction Box Housing Remote Mount Display (white) 24 AWG (blue) 24 AWG (black) 22 AWG (red) 22 AWG 4-20 mA NOTE Wire colors provided above are per Belden 3084A DeviceNet cable. Wire color may vary depending on cable selected.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 As standard, the 3051S Quick Connect arrives properly assembled to the Quick Connect Wiring SuperModule and is ready for installation. Cordsets and Field Wireable Connectors (in shaded area) are sold separately.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 For 3051S transmitters with conduit electrical connectors GE or GM, refer to Conduit Electrical the cordset manufacturer’s installation instructions for wiring details. For FM Connector Wiring Intrinsically Safe, non-incendive or FM FISCO Intrinsically Safe hazardous...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Use the following procedure and Figure 2-13 to install the LCD display: 1. IF the transmitter is installed in a loop, THEN secure the loop and disconnect power. 2. Remove the transmitter cover opposite the field terminal side. Do not remove the instrument covers in explosive environments when the circuit is live.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 To install a 305 Integral Manifold to a 3051S transmitter: Rosemount 305 Integral Manifold Installation 1. Inspect the PTFE SuperModule O-rings. If the O-rings are undamaged, Procedure reusing them is recommended. If the O-rings are damaged (if they have nicks or cuts, for example), replace them with new O-rings.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 7. Subtract the number of threads showing (after tightening) from the total threads to calculate the revolutions engaged. Further tighten until a minimum of 3 rotations is achieved. 8. For block and bleed manifold, verify the bleed screw is installed and tightened.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 To install a 305 Integral Manifold to a 3051S MultiVariable transmitter: Rosemount 305 Integral Manifold Installation 1. Inspect the PTFE SuperModule O-rings. If the O-rings are undamaged, Procedure reusing them is recommended. If the O-rings are damaged (if they have nicks or cuts, for example), replace them with new O-rings.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Improper installation or operation of manifolds may result in process leaks, Manifold Operation which may cause death or serious injury. Always perform a zero trim on the transmitter/manifold assembly after installation to eliminate any shift due to mounting effects.
Page 41
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 3. After zeroing the transmitter, close the equalizing valve. Drain/ Drain/ Vent Vent Valve Valve Equalize (closed) Isolate Isolate (closed) (open) Process 4. Open the block valve on the low pressure side of the transmitter to return the transmitter to service.
Page 42
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Five-valve Natural Gas configurations shown: In normal operation, the two block valves between the process and instrument ports will be open, and the equalizing valves will be closed. Test Test...
Page 43
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 3. Open the equalize valve on the low pressure (downstream) side of the transmitter. The manifold is now in the proper configuration for zeroing the transmitter. Test Test (Plugged) (Plugged)
Page 44
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 2-30...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Commissioning consists of testing the transmitter and verifying transmitter COMMISSIONING ON configuration data. 3051S transmitters can be commissioned either before or THE BENCH WITH HART after installation. Commissioning the transmitter on the bench before installation using a 375 Field Communicator or AMS ensures that all transmitter components are in working order.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Wiring Diagrams Bench Hook-up Connect the bench equipment as shown in Figures 3-1 and 3-2, and turn on the Field Communicator by pressing the ON/OFF key or log into AMS. The Field Communicator or AMS will search for a HART-compatible device and indicate when the connection is made.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 REVIEW NOTE Information and procedures in this section that make use of Field CONFIGURATION DATA Communicator fast key sequences and AMS assume that the transmitter and communication equipment are connected, powered, and operating correctly.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 (Version 3.3) FIELD COMMUNICATOR Field Communicator User Inferface Figure 3-3. Traditional Interface The corresponding Menu Tree can be viewed on page 3-6. The fast key sequence can be viewed on page 3-7.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Traditional Interface Menu Tree 1. PROCESS 1. Pressure 2. Percent of Range VARIABLES 3. Analog Output 4. Sensor Temperature 5. Scaled Variable 6. Process Variable is 1. Self test 2. DIAG/SERVICE 2.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Device Dashboard Menu Tree Figure 3-5. 3051S Device Dashboard - Overview Home Identification Overview 1 Overview 1 Tag 1 Device Status 2 Configure 2 Model 2 Comm Status 3 Service Tools...
Page 52
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 3-6. 3051S Device Dashboard - Configure Tag, Description, Message, Date, Pressure Units, Temperature Units, Transfer Function, URV, LRV Basic Setup Home Configure Guided Setup 1 Overview 1 Guided Setup...
Page 53
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 3-7. 3051S Device Dashboard - Service Tools Home Service Tools Device Alerts 1 Overview 1 Device Alerts 1 Refresh Alerts 2 Configure 2 Variables Variables 2 Configuration Changed 3 Service Tools...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The following menu indicates fast key sequences for common functions. A Traditional Fast Key check ( ) indicates the basic configuration parameters. At minimum, these Sequence parameters should be verified as part of the configuration and startup procedure.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The following menu indicates fast key sequences for common functions. A Device Dashboard Fast check ( ) indicates the basic configuration parameters. At minimum, these Key Sequence parameters should be verified as part of the configuration and startup procedure.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Before performing other transmitter on-line operations, review the digital CHECK OUTPUT output parameters to ensure that the transmitter is operating properly and is configured to the appropriate process variables. Process Variables The process variables for the 3051S provide transmitter output, and are continuously updated.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 BASIC SETUP The PV Unit command sets the process variable units to allow you to monitor Set Process Variable your process using the appropriate units of measure. Units 1, 3, 2 Traditional Fast Keys Field Communicator v3.3...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 3-8. Square Root Output Transition Point Sq. Root Full Scale Full Scale Curve Flow (%) Output (mA dc) Sq. Root Curve Transition Point Transition Point Linear Section Slope=42 Slope=1...
Page 59
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 NOTE If the transmitter security jumper/switch is ON, adjustments to the zero and span will not be able to be made. Refer to “Configure Security and Alarm” on page 2-13 for security information.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Rerange with a Pressure Input Source and the Local Zero and Span buttons (option D1) Reranging using the local zero and span adjustments and a pressure source is a way of reranging the transmitter.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Field Communicator v3.3 Enter the fast key sequence “Damping.” AMS v7.0 Right click on the device and select “Configure” from the menu. 1. In the “Basic Setup” tab, enter the damping value in the “Damp” field, click Apply.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 DETAILED SETUP 3051S transmitters automatically and continuously perform self-diagnostic Failure Mode Alarm and routines. If the self-diagnostic routines detect a failure, the transmitter drives Saturation the output to configured alarm values. The transmitter will also drive the output to configured saturation values if the applied pressure goes outside the 4-20 mA range values.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 To configure alarm and saturation levels with a Field Communicator or AMS Alarm and Saturation perform the following procedure: Level Configuration 1, 4, 2, 7 Traditional Fast Keys Field Communicator v3.3...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Transmitters set to multidrop mode handle saturation and alarm conditions Alarm and Saturation differently. Values for Multidrop Mode Alarm Conditions: • Primary variable is sent with a status bit set •...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 AMS v7.0 Right click on the device and select “Configure” from the menu 1. In the “Analog Output” tab, locate the “Configuration Pressure Alerts” box, enter “Press Hi Alert Val” and “Press Lo Alert Val” to configure the pressure alerts.
Page 66
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Field Communicator v3.3 To configure the scaled variable with a Field Communicator, perform the following procedure: 1. From the HOME screen follow the fast key sequence “Scaled Variable Configuration.” 2. Select OK after the control loop is set to manual.
Page 67
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 AMS v7.0 Right click on the device and select “Device Configuration” then select “SV Config” from the menu 1. Click Next after setting the control loop to manual. 2. Enter desired scaled variable units in “Enter SV units” box and click Next.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 DP Flow Example A differential transmitter is used in conjunction with an orifice plate in a flow application where the differential pressure at full scale flow is 125 inH2O. In this particular application, the flow rate at full scale flow is 20,000 gallons of water per hour.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The Loop Test command verifies the output of the transmitter, the integrity of Loop Test the loop, and the operations of any recorders or similar devices installed in the 1, 2, 2 Traditional Fast Keys loop.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 5. Check the reference meter installed in the test loop to verify that it displays the commanded output value. a. If the values match, the transmitter and the loop are configured and functioning properly.
Page 72
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 5. Locate the saved transmitter configuration file. a. Select “Offline” b. Select “Saved Configuration” c. Select either “Internal Flash Contents” or “Configuration EM Contents” depending on where the configuration was stored per step 2b.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 AMS v7.0 Applying a User Configuration Any amount of user configurations can be created for the application. They can also be saved, and applied to connected devices or to devices in the Device List or Plant Database.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 AMS v7.0 Right click on the device and select “Configure” from the menu 1. In the “HART” tab, use the drop down menu to select “Burst Mode ON or OFF.” For “Burst option” select the desired properties from the drop down menu.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The 3051S is set to address zero (0) at the factory, which allows operation in the standard point-to-point manner with a 4–20 mA output signal. To activate multidrop communication, the transmitter address must be changed to a number from 1 to 15.
Page 76
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 3-32...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Section 4 Operation and Maintenance Overview ........page 4-1 Calibration for HART Protocol .
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table 4-1. Recommended Calibration Tasks Transmitter Bench Calibration Tasks Field Calibration Tasks 1. Set output configuration parameters: 1. Reconfigure parameters if necessary. 3051S_ CD 3051S_ CG a. Set the range points.
Page 79
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Calibrate the sensor • Sensor Trim (page 4-6) • Zero Trim (page 4-6) Calibrate the 4–20 mA output • 4–20 mA Output Trim (page 4-8); or • 4–20 mA Output Trim Using Other Scale (page 4-9) Figure 4-1 on page 4-3 illustrates 3051S transmitter data flow.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Calibration frequency can vary greatly depending on the application, Determining Calibration performance requirements, and process conditions. Use the following Frequency procedure to determine calibration frequency that meets the needs of your application.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 To decide which trim procedure to use, you must first determine whether the Choosing a Trim analog-to-digital section or the digital-to-analog section of the transmitter Procedure electronics needs trimming. Refer to Figure 4-1 and perform the following procedure: 1.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Zero Trim NOTE The transmitter must be within three percent of true zero (zero-based) in order 1, 2, 3, 3, 1 Fast Keys to calibrate with zero trim function. 3, 4, 1, 3...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 NOTE Select pressure input values so that lower and upper values are equal to or outside the 4 and 20 mA points. Do not attempt to obtain reverse output by reversing the high and low points.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Digital-to-Analog Trim Field Communicator 1, 2, 3, 2, 1 Fast Keys To perform a digital-to-analog trim with a Field Communicator, perform the following procedure. 3, 4, 2 Device Dashboard Fast Keys 1.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The Scaled D/A Trim command matches the 4 and 20 mA points to a user Digital-to-Analog Trim selectable reference scale other than 4 and 20 mA (for example, 1 to 5 volts if...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The Recall Factory Trim—Analog Output command allows the restoration of Recall Factory Trim— the as-shipped factory settings of the analog output trim. This command can Analog Output be useful for recovering from an inadvertent trim, incorrect Plant Standard or...
Page 87
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The following specifications show the static pressure effect for the 3051S Range 4 and Range 5 transmitters used in differential pressure applications: Zero Effect: ± 0.1% of the upper range limit per 1000 psi (69 bar) for line pressures from 0 to 2000 psi (0 to 138 bar) For line pressures above 2000 psi (138 bar), the zero effect error is ±...
Page 88
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 In this example: URV = 1500 inH O (3.74 bar) LRV = 500 inH O (1.25 bar) 1200 psi (82.74 bar) ± 0.01/1000 To calculate the low trim (LT) value: LT = 500 + (0.01/1000)(500)(1200)
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 In addition to output, the LCD display displays abbreviated operation, error, Diagnostic Messages and warning messages for troubleshooting. Messages appear according to their priority; normal operating messages appear last. To determine the cause of a message, use a Field Communicator or AMS to further interrogate the transmitter.
Page 90
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 XMRT INFO A non-volatile memory fault has been detected in the transmitter memory by the memory verification routine. The memory fault is in a location containing transmitter information. To correct this problem, use a Field Communicator or AMS to interrogate and reconfigure the appropriate portion of the transmitter memory.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 FIELD UPGRADES Each housing and each SuperModule is labeled individually, so it is Labeling imperative that the approval codes on each label match exactly during upgrade. The label on the SuperModule reflects the replacement model code for reordering an assembled unit.
Page 92
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 4-16...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Section 5 Troubleshooting Overview ........page 5-1 Safety Messages .
Page 94
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table 5-1. Rosemount 3051S troubleshooting table Symptom Corrective Actions Transmitter milliamp reading is zero Verify power is applied to signal terminals Check power wires for reversed polarity Verify terminal voltage is 10.5 to 42.4 V dc...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Do not remove the instrument cover in explosive atmospheres when the DISASSEMBLY circuit is live. PROCEDURES Be aware of the following: Remove from Service • Follow all plant safety rules and procedures.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The Standard Interface Assembly, Adjustment Interface Assembly, Safety Remove Interface Certified Electronics Assembly (with yellow casing), or HART Diagnostics Assembly Electronics Assembly (black casing with white label) is located in the compartment opposite the terminal side in the PlantWeb housing.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 REASSEMBLY IMPORTANT The V-Seal must be installed at the bottom of the housing. PROCEDURES Attach SuperModule to 1. Apply a light coat of low temperature silicon grease to the SuperModule threads and O-ring.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 1. Inspect the SuperModule PTFE O-rings. If the O-rings are undamaged, Reassemble the Process reusing them is recommended. If the O-rings are damaged (if they have Flange nicks or cuts, for example), replace them with new O-rings.
Page 99
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table 5-2. Bolt Installation Torque Values Bolt Material Initial Torque Value Final Torque Value CS-ASTM-A445 Standard 300 in-lb. (34 N-m) 650 in-lb. (73 N-m) 316 SST—Option L4 150 in-lb. (17 N-m) 300 in-lb.
Page 100
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The 3051S safety certified pressure transmitter is certified to: CERTIFICATION Low Demand; Type B SIL 2 Capability for hardware (single use transmitter) SIL 3 Capability for software (multiple use transmitter)
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 To commission the 3051S Safety Certified Transmitter, use the HART “Menu COMMISSIONING Tree” on page 3-5 and “Fast Key Sequence” on page 3-6. NOTE Transmitter output is not safety-rated during the following: configuration changes, multidrop, and loop test.
Page 104
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Switches installed 1. If using a communicator, use the following fast key sequence to set the Alarm and Saturation values. Alarm Levels - Fast Key; 1, 4, 2, 7, 7 Saturation Levels - Fast Key;...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 OPERATION AND MAINTENANCE The following proof tests are recommended. Proof Test Proof test results and corrective actions taken must be documented at http://rosemount.d1asia.ph/rosemount/safety/ReportAFailure_newweb.asp (Report a Failure button) in the event that an error is found in the safety functionality.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Inspection Visual Inspection Not required Special Tools Not required Product Repair The 3051S is repairable by major component replacement. All failures detected by the transmitter diagnostics or by the proof-test must be reported.
The Rosemount 3051S HART Diagnostic transmitter is available with the new Enhanced EDDL interface; DD revision 3051S HDT Dev. 1 Rev. 1 is required. USER INTERFACE The following screen shots are taken from Emerson Process Management’s ™ Device Manager, version 7.6 or higher. Screens from other hosts that support Enhanced EDDL will look very similar.
Rosemount 3051S Series August 2010 STATISTICAL PROCESS MONITORING Emerson Process Management has developed a unique technology, SPM Introduction Statistical Process Monitoring (SPM), that provides a means for early detection of abnormal situations in a process environment. The technology is based on the premise that virtually all dynamic processes have a unique noise or variation signature when operating normally.
(or average) is calculated on the unfiltered pressure signal, the standard deviation calculated from the filtered pressure signal. These statistical values are available via HART and handheld communication devices like the 375 Field Communicator or asset management software like Emerson Process ™ Management’s AMS Device Manager.
Page 111
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 SPM also contains a learning module that establishes the baseline values for the process. Baseline values are established under user control at conditions considered normal for the process and installation. These baseline values are made available to a decision module that compares the baseline values to the most current values of the mean and standard deviation.
Page 112
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Further detail of the operation of the SPM diagnostic is shown in the Figure 7-4 flowchart. This is a simplified version showing operation using the default values. While SPM continuously calculates the mean and standard deviation values, the learning and decision modules must be turned on to operate.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The mean and standard deviation values can be made available to other Assigning Statistical systems via HART communications or devices that convert HART variables to Values to Outputs analog 4 – 20 mA outputs, such as Rosemount’s 333 Tri-Loop. Either value can be assigned to be a secondary or tertiary variable.
Page 114
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 7-6 is the entry screen for the Statistical Process Monitoring diagnostic. This screen is the main screen for the SPM diagnostic results and status. The process for correct operation of the SPM diagnostic is: •...
Page 115
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Power Interruption This is used to direct what the diagnostic should do in the case of a power interruption or if the diagnostic is manually disabled and then enabled. The...
Page 116
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 7-10. SPM Status screen showing the sensitivity settings Table 7-1. Standard Deviation Sensitivity choices (regardless of transmitter type) An 80% change from the baseline value will trip the diagnostic.
Page 117
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 7-11. Sensitivity settings description High Figure 7-11 Illustrates how the sensitivity selection affects the diagnostic trip values. The second configuration setting is the action the transmitter is to take if the diagnostic trips.
Page 118
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table 7-3. SPM Diagnostic Configuration defaults Parameter Default Learning/Monitoring Period 3 minutes High Pass Filter Power Interruption Monitor Variation Change Sensitivity Medium Variation Change Action Alert Mean Change Sensitivity Medium...
Page 119
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Standard Deviation and Mean Verification Criteria The Standard Deviation and Mean configuration parameters define the maximum deviation from the original Baseline values derived during the Learning/Verification Mode. If these maximum values are exceeded, the diagnostic will not start/go to the Monitoring Mode, but will continue Verifying.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 7-13. Operational The Operational Values screen contains the parameter values used in the Values Screen SPM diagnostic in tabular form. SPM Operation Figure 7-14. Turning on the SPM diagnostic...
Page 121
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Turning ON the SPM Diagnostic The SPM diagnostic is enabled by selecting ON in the “Control” section shown on Figure 7-14. Upon selecting ON in the “Control” section, the diagnostic will automatically begin “Learning” with the following exception: if valid baseline values have been previously established and “Monitor”...
Page 122
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table 7-7. Possible causes of SPM diagnostic events Diagnostic Condition Potential Cause Corrective Action Plugged impulse line (DP Follow facility procedure to check for and clear only) plugged impulse lines. Both lines must be checked as the SPM diagnostic cannot determine if the plug is on the high or low side.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Resetting SPM The two options for resetting the SPM diagnostic are: Reset When Reset is selected the system will ask for confirmation and then will clear the status indicators in the SPM and reset the “Time Since Detection”...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 VARIABLE LOGGING The 3051S HART Diagnostic transmitter’s Variable Logging can be used in a Overview number of ways. Figure 7-15 shows the Pressure Event Logging screen. Figure 7-16 shows the Temperature Event Logging screen. On both of these screens there are two basics sets of functionality.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Time Outside Sensor Limits gives the operator/maintenance personnel an indication of possible misapplication of the transmitter. The Lower and Upper operate the same. They both include a Time Since 1st Event and Total time.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 7-16. Temperature Event Logging screen Time Outside Sensor Limits gives the operator/maintenance personnel an indication of possible misapplication of the transmitter. The Lower and Upper operate the same. They both include a Time Since 1st Event and Total time.
Page 127
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 7-17. Process Pressure Alerts screen Pressure Alert Setting This is the configuration section for the Pressure Process Alert. Selecting ON enables the diagnostic. The High Alert Value and the Low Alert Value are the trip values for the diagnostic.
Page 128
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure 7-18. Temperature Process Alerts screen Temperature Alert Setting This is the configuration section for the Temperature Process Alert. Selecting ON enables the diagnostic. The High Alert Value and the Low Alert Value are the trip values for the diagnostic.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 ROSEMOUNT 333 HART TRI-LOOP CONFIGURATION The Rosemount 333 HART Tri-Loop can be used in conjunction with the Overview Rosemount 3051S with Advanced HART Diagnostics to acquire up to two more variables via 4-20mA analog signals. The additional two outputs are selected by the user and can include: Pressure, Temperature, Scaled Variable, Standard Deviation, or Mean.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The 3051S safety certified pressure transmitter is certified to: 3051S SAFETY Low Demand; Type B INSTRUMENTED SIL 2 Capability for hardware (single use transmitter) SYSTEMS SIL 3 Capability for software (multiple use transmitter)
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Use any HART-compliant master to communicate with and verify COMMISSIONING configuration of the 3051S Safety Certified transmitter with Advanced HART Diagnostics. Use the HART "“Menu Tree” on page 3-5 and “Fast Key Sequence”...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Rosemount Alarm Level (3)(4) Custom Alarm Level Normal Operation 3.6 - 3.8 mA 4 mA 20 mA 20.2 - 23.0 3.7 - 3.9 mA 20.1 - 20.5 mA low saturation high saturation (1) Transmitter Failure, hardware or software alarm in LO position.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 1. Bypass the safety function and take appropriate action to avoid a false trip. 2. On Field Communicator enter the Fast Key Sequence 3, 5, 1. 3. Select “4 Other.”...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The 3051S must be operated in accordance to the functional and SPECIFICATIONS performance specifications provided in the 3051S Product Data Sheet (document number 00813-0100-4801). Failure Rate Data The FMEDA report includes failure rates and common cause Beta factor estimates.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Appendix A Specifications and Reference Data Performance Specifications ..... . . page A-1 Functional Specifications .
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Transmitter with In-Line Sensor Module Absolute Pressure (3051S_TA) Gage Pressure (3051S_TG) Ultra Classic Ranges 1 - 4 ±0.025% of span ±0.055% of span For spans less than 10:1, For spans less than 10:1, ±[0.004(URL / Span)]% of span...
(1) Warranty details can be found in Emerson Process Management Terms & Conditions of Sale, Document 63445, Rev G (10/06). (2) Rosemount Ultra and Ultra for Flow transmitters have a limited warranty of twelve (12) years from date of shipment. All other provisions of Emerson Process Management standard limited warranty remain the same.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Ambient Temperature Effect Transmitter with Coplanar Sensor Module (Single Variable) Differential Pressure: (3051S_CD) Gage Pressure: (3051S_CG) Ultra Classic Ultra for Flow per 50 °F (28 °C) per 50 °F (28 °C) -40 to 185 °F (-40 to 85 °C)
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Line Pressure Effect 3051S_CD Ultra and Ultra for Flow Classic Zero Error Range 2-3 ± 0.025% URL per 1000 psi (69 bar) ± 0.05% URL per 1000 psi (69 bar) Range 0 ±...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 FUNCTIONAL SPECIFICATIONS Range and Sensor Limits Transmitter with Coplanar Sensor Module (Single Variable) DP Sensor GP Sensor AP Sensor (3051S_CD, 3051S_LD) (3051S_CG, 3051S_LG) (3051S_CA, 3051S_LA) Range Lower (LRL) Upper (URL)
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Minimum Span Limits Transmitter with Coplanar Sensor Module (Single Variable) DP Sensor GP Sensor AP Sensor (3051S_CD, 3051S_LD) (3051S_CG, 3051S_LG) (3051S_CA, 3051S_LA) Ultra & Range Ultra for Flow Classic Ultra...
Page 142
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Load Limitations Maximum loop resistance is determined by the voltage level of the external power supply, as described by: 3051S Maximum Loop Resistance = 43.5 * (Power Supply Voltage – 10.5)
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Temperature Limits Ambient -40 to 185 °F (-40 to 85 °C) With LCD display : -40 to 175 °F (-40 to 80 °C) With option code P0: -20 to 185 °F (-29 to 85 °C) (1) LCD display may not be readable and LCD updates will be slower at temperatures below -4 °F (-20 °C).
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Humidity Limits 0–100% relative humidity Turn-On Time When power is applied to the transmitter during startup, performance will be within specifications per the time period described below: Transmitter Turn-On Time (Typical)
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 3051S_L Mounting Flange Zinc-cobalt plated CS or 316 SST 3051S_L Seal Extension CF-3M (Cast 316L SST, material per ASTM A743) or CW-12MW (Cast C-276, material per ASTM A494) Non-Wetted Parts...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 DIMENSIONAL DRAWINGS Figure A-1. Transmitter with Coplanar Sensor Module and Flange PlantWeb Housing Junction Box Housing Wireless Housing Front View Front View Front View Extended Range 4.20 (107) Antenna 4.20 (107) 90°...
Page 151
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure A-2. Transmitter with Coplanar Sensor Module and Traditional Flange PlantWeb Housing Junction Box Housing Wireless Housing Front View Front View Front View Extended Range Antenna 90° Long Range 6.93...
Page 152
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure A-3. Transmitter with In-Line Sensor Module PlantWeb Housing Junction Box Housing Wireless Housing Front View Front View Front View Extended Range Antenna 4.20 (107) 4.20 (107) Long 90° 6.93...
Page 153
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure A-4. Coplanar Mounting Configurations (B4 Bracket) Pipe Mount Panel Mount Front View Front View Front View 2.58 (66) 4.55 (116) 6.15 (156) 2.81 (71) 4.70 (120) 3.54 (90) 6.25 (159)
Page 154
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure A-5. Traditional Mounting Configurations Pipe Mount Pipe Mount (Flat Bracket) Panel Mount 10.70 (272) 2.62 (67) 8.18 (196) 13.03 (208) (331) 0.93 (24) 5.32 4.85 (135) (123) 3.42 (87) 2.62...
Page 155
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Figure A-7. Remote Display Mounting Configurations (B4 Bracket) Pipe Mount Panel Mount Front View Side View 2.33 5.19 (132) (59) 1.82 (46) 4.48 (114 2.66 2.66 (68) (68) 6.90 (175) 6.15...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 ORDERING INFORMATION Table A-1. Rosemount 3051S Scalable Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery.
Page 158
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-1. Rosemount 3051S Scalable Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 159
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-1. Rosemount 3051S Scalable Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 160
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-1. Rosemount 3051S Scalable Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 161
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-1. Rosemount 3051S Scalable Coplanar Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 162
(19) Not available with housing style codes 00, 2E, 2F, 2G, 2M, 5A, 5J, or 7J. (20) Requires PlantWeb housing and Hardware Adjustments option code D1. Limited availability depending on transmitter type and range. Contact an Emerson Process Management representative for additional information.
Page 163
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-2. Rosemount 3051S Scalable In-Line Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 164
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-2. Rosemount 3051S Scalable In-Line Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 165
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-2. Rosemount 3051S Scalable In-Line Pressure Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 166
(13) Not available with housing style codes 00, 01, 2E, 2F, 2G, 2M, 5A, 5J, or 7J. (14) Requires PlantWeb housing and Hardware Adjustments option code D1. Limited availability depending on transmitter type and range. Contact an Emerson Process Management representative for additional information.
Page 167
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-3. Rosemount 3051S Liquid Level Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 168
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-3. Rosemount 3051S Liquid Level Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 169
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-3. Rosemount 3051S Liquid Level Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 170
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-3. Rosemount 3051S Liquid Level Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 171
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Table A-3. Rosemount 3051S Liquid Level Transmitter Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 172
(13) Not available with housing style codes 00, 2E, 2F, 2G, 2M, 5A, 5J, or 7J. (14) Requires PlantWeb housing and Hardware Adjustments option code D1. Limited availability depending on transmitter type and range. Contact an Emerson Process Management representative for additional information.
Page 173
Rosemount 3051S Series August 2010 Table A-4. Housing Kit for Rosemount 3051S Series Ordering Information ★ The Standard offering represents the most common options. The starred options (★) should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 174
(5) Not available with output code F. (6) Not available with Housing code 7J. (7) See the 3051S Reference Manual (document number 00809-0100-4801) for cable requirements. Contact an Emerson Process Management representative for additional information. (8) Not available with output code F, option code DA1, or option code DA2. Only available on Housing Style codes 3A, 3B, 3C, or 3J.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 EXPLODED VIEW DIAGRAM The following drawing shows the name and location for commonly ordered spare parts. PlantWeb Housing Terminal Block Cover O-ring Transmitter Electronics Cover Module O-Ring Coplanar Flange Process Flange O-Ring...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 SPARE PARTS See Rosemount 3051S_C, 3051S_T & 3051S_L ordering tables in Appendix A (A-23, A-29, and A-33 respectively) for ordering spare sensor modules. - Typical Model Number 3051S1CD2A2000A00 Electronics Board Assembly Hardware (PlantWeb® Housing)
Page 177
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Covers Aluminum Electronics Cover; Cover and O-ring 03151-9030-0001 316L SST Electronics Cover; Cover and O-ring 03151-9030-0002 Housing Miscellaneous External Ground Screw Assembly (Option D4): Screw, clamp, 03151-9060-0001 washer Housing V-Seal for both PlantWeb and Junction Box housings...
Page 178
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Flange Adapter Union Part Number Nickel-plated Carbon Steel 03151-9259-0005 316 SST 03151-9259-0002 Cast C-276 03151-9259-0003 Cast Alloy 400 03151-9259-0004 Drain/Vent Valve Kits (each kit contains parts for one transmitter) Part Number...
Page 179
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Flange/Adapter Bolt Kit {73 mm (2.88 in.)} Carbon Steel (set of 4) 03151-9281-0001 316 SST (set of 4) 03151-9281-0002 ANSI/ASTM-A-193-B7M (set of 4) 03151-9281-0003 Alloy K-500 (set of 4) 03151-9281-0004 Manifold/Flange Kit {57 mm (2.25 in.)}...
Page 180
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 A-46...
Administration (OSHA). European Directive Information The EC declaration of conformity for all applicable European directives for this product can be found at www.rosemount.com. A hard copy may be obtained by contacting an Emerson Process Management representative. ATEX Directive (94/9/EC) Emerson Process Management complies with the ATEX Directive.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 HAZARDOUS North American Certifications LOCATIONS FM Approvals CERTIFICATIONS Explosion-proof for Class I, Division 1, Groups B, C, and D, T5 (T = 85 °C); Dust Ignition-proof for Class II and Class III, Division 1, Groups E, F, and G, T5 (T = 85 °C);...
Page 183
3. The 3051S does not comply with the requirements of EN 60079-1 Clause 5.2, Table 2 for all joints. Contact Emerson Process Management for information on the dimensions of flameproof joints.
Page 184
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Japanese Certifications TIIS Flameproof Ex d IIC T6 Certificate Description TC15682 Coplanar with Junction Box Housing TC15683 Coplanar with PlantWeb Housing TC15684 Coplanar with PlantWeb Housing and LCD Display TC15685...
Page 185
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 11. During installation, use, and maintenance of the pressure transmitter, observe the following standards: a. GB3836.13-1997 “Electrical apparatus for explosive gas atmospheres Part 13: Repair and overhaul for apparatus used in explosive gas atmospheres”...
Page 186
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 8. During installation, use, and maintenance of the pressure transmitter, observe the following standards: a. GB3836.13-1997 “Electrical apparatus for explosive gas atmospheres Part 13: Repair and overhaul for apparatus used in explosive gas atmospheres”...
Page 187
3. The 3051S does not comply with the requirements of IEC 60079-1 Clause 5.2, Table 2 for all joints. Contact Emerson Process Management for information on the dimensions of flameproof joints.
Page 188
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 IECEx Dust Certificate No. IECExBAS09.0014X Ex tD A20 T105°C (-20°C ≤ Tamb ≤ 85°C) Vmax = 42.4 V A = 22 mA IP66 Special conditions for safe use (x) 1. Cable entries must be used which maintain the ingress protection of the enclosure to at least IP66.
Page 189
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 IECEx Type n Certificate No.: IECExBAS04.0018X Ex nC IIC T4 (T = -40 °C to 70 °C) Ui = 45 Vdc MAX IP66 Special conditions for safe use (x) The apparatus is not capable of withstanding the 500 V insulation test required by Clause 8 of IEC 60079-15.
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 EUROPEAN ATEX DIRECTIVE INFORMATION Rosemount 3051S pressure transmitters that have the following label CENELEC/BASEEFA attached, have been certified to comply with Directive 94/9/EC of the European Parliament and the Council as published in the Official Journal of the European Communities No.
Page 224
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Type n housing label • Marking for type n protection: Ex nAnL IIC T4 (-40 °C 70 °C) II 3 G amb Ui = 45 Vdc MAX BASEEFA ATEX certificate number: BAS01ATEX3304X Dust housing label •...
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 The Rosemount 3051S and 300S pressure transmitters that have the CENELEC/KEMA following label attached, have been certified to comply with Directive 94/9/EC FLAMEPROOF of the European Parliament and the Council as published in the Official Journal of the European Communities No.
Page 226
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 B-46...
Page 227
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 Index Configuration . . . 3-19 Alarm and saturation Address Failure mode alarm and saturation Applying a user configuration ... 3-31 .
Page 228
Reference Manual 00809-0100-4801, Rev EA Rosemount 3051S Series August 2010 ... . 3-24 ..3-13 Re-mapping Transfer function ..2-18...
Page 232
PlantWeb is a mark of one of the Emerson Process Management companies. HART is a registered trademark of the HART Communications Foundation. ASP Diagnostics Suite is a trademark of one of the Emerson Process Management companies. Syltherm and D.C. are registered trademarks of Dow Corning Co.