Appendix A Specifications And Reference Data; Performance Specifications - Emerson Rosemount 1208A Reference Manual

Level and flow transmitter
Hide thumbs Also See for Rosemount 1208A:
Table of Contents

Advertisement

Reference Manual
00809-0100-7062
A
Specifications and reference data
A.1

Performance specifications

A.1.1
General
Reference conditions
Measurement target: Stationary metal plate, no disturbing objects
Temperature: 59 to 77 °F (15 to 25 °C)
Ambient pressure: 14 to 15 psi (960 to1060 mbar)
Relative humidity: 25-75%
Damping: Default value, 2 s
Instrument accuracy (under reference conditions)
±0.08 in. (±2 mm)
Repeatability
±0.04 in. (±1 mm)
Ambient temperature effect
±0.04 in. (±1 mm)/10 K
Sensor update rate
Minimum 1 update per second (typically 5 updates per second)
Maximum level rate
200 mm/s
A.1.2
Measuring range
Maximum measuring range
49 ft. (15 m)
(1)
Refers to inaccuracy according to IEC 60770-1 when excluding installation dependent offset. See the IEC 60770-1 standard
for a definition of radar specific performance parameters and if applicable corresponding test procedures.
(2)
The measuring range is limited to 33 ft. (10 m) for oil-based media (dielectric constant < 10). Also note that a combination
of adverse process conditions, such as heavy turbulence, foam, and condensation, together with products with poor
reflection may affect the measuring range.
Reference Manual
(1)
(2)
Specifications and reference data
October 2022
69