Reference Manual
00809-0200-4480
A
Specifications and reference data
A.1
Performance specifications
A.1.1
General
Reference conditions
•
Measurement target: Stationary metal plate, no disturbing objects
•
Temperature: 59 to 77 °F (15 to 25 °C)
•
Ambient pressure: 14 to 15 psi (960 to1060 mbar)
•
Relative humidity: 25-75%
•
Damping: Default value, 2 s
Instrument accuracy (under reference conditions)
±0.08 in. (±2 mm)
Repeatability
±0.04 in. (±1 mm)
Ambient temperature effect
±0.04 in. (±1 mm)/10 K
Sensor update rate
Minimum 1 update per second (typically 5 updates per second)
Maximum level rate
200 mm/s
A.1.2
Measuring range
Maximum measuring range
33 ft. (10 m)
(1)
Refers to inaccuracy according to IEC 60770-1 when excluding installation dependent offset. See the IEC 60770-1 standard
for a definition of radar specific performance parameters and if applicable corresponding test procedures.
(2)
Note that a combination of adverse process conditions, such as heavy turbulence, foam, and condensation, together with
products with poor reflection may affect the measuring range.
Reference Manual
(1)
(2)
Specifications and reference data
May 2021
65