11 Preventive maintenance

11.5.2 Milling zero point and B-axis calibration body

90
Planmeca PlanMill 50 S
7. Place the new workpiece in the workpiece holder.
If not using a new blank workpiece, ensure that the workpiece is
positioned and oriented correctly so that the milling is made on blank
space of the workpiece.
8. Insert the clamping ring and turn in an anti-clockwise direction up to the
stop.
9. Tighten the locking screws on the clamping ring hand tight.
10. Check the tightness of the workpiece.
11. Close the lid.
12. Move the workpiece holder to home position by pressing the move to
home position button:
1. Attach work piece, as described in section "Attaching workpiece" on
page 88, if not yet attached.
2. Switch on the suction.
3. On the desktop you will find a Service shortcut. Open the folder by
double clicking the Service icon.
User's manual