7 Adapters
7.2.2.2 Preparing adapter for test milling
38
Planmeca PlanMill 50 S
No damages in the bearing surface
The milling unit and its zero-point must be set correctly
Always clamp the adapter in the same position
Screw connections
The 6 screws should always be tightened in the same order for a
reproducible positioning
Only screw the 6 countersunk screws loosely at first
Afterwards tighten the screws crosswise (always in the same
order)
NOTE
Carefully follow the directions on general cleanliness and handling.
The two test pins, available from your local premilled blank supplier, are
thoroughly inserted into the middle positions (2 + 5), pushed in to the limit,
and firmly clamped with the clamping screw by using a torque wrench.
Check that both test pins are attached directly without a gap to the
supporting surface.