7.2.1.1 General cleanliness and handling
7.2.1.2 Preparing adapter for test milling
User's manual
The adapter has to be calibrated in Planmeca PlanMill 50 S milling units
before the first use to ensure an optimal positioning of the DESS system.
Therefore the following aspects must be considered:
Cleanliness of the blank adapter (milling unit) and of the DESS adapter
General Cleanliness
No damages in the bearing surface
The milling unit and its zero-point must be set correctly
Always clamp the adapter in the same position
Screw connections
The 6 screws should always be tightened in the same order for a
reproducible positioning
Only screw the 6 countersunk screws loosely at first
Afterwards tighten the screws crosswise (always in the same
order)
NOTE
Please follow directions on general cleanliness and handling.
The two test pins, available from your local premilled blank supplier, are
thoroughly inserted into the middle positions (2 + 5), pushed in to the
limit, and firmly clamped with the clamping screw by using a torque
wrench.
The DESS adapter is then fixed into the clean milling unit reception. The
screws should always be screwed in loosely at first and tightened
afterwards in the same order, 1 - 6.
The test pins are milled directly on the milling unit with the scripted test
program available from your local dealer. A 2mm metal tool (T2) is
required for the milling process.