Adjustment & Calibration; Required Tools - Planmeca Proline EC pan Technical Manual

Table of Contents

Advertisement

Chapter
D
NOTE
This chapter contains the information required to set up and calibrate the Planmeca
Proline EC Panoramic X-ray. In case your unit is equipped with Dimax digital sensor
head, perform the adjustments according to the instructions given in chapter "DIMAX
DIGITAL SYSTEM ADJUSTMENT" on page E-1.
Protect yourself from radiation when you are checking the beam alignment. Since radiation
safety requirements vary from state to state and country to country it is the responsibility of
the installer to ensure that the correct precautions are observed.
The display values shown in this guide are only examples and should not be interpreted as
recommended values unless otherwise stated.
WARNING
IT IS VERY IMPORTANT THAT THE ROOM IN WHICH THE X-RAY IS INSTALLED AND
THE POSITION FROM WHICH THE USER OPERATES THE EQUIPMENT ARE COR-
RECTLY SHIELDED. SINCE RADIATION SAFETY REQUIREMENTS VARY FROM COUN-
TRY TO COUNTRY AND STATE TO STATE IT IS THE RESPONSIBILITY OF THE
INSTALLER TO ENSURE THAT ALL LOCAL SAFETY REGULATIONS ARE MET.

1 REQUIRED TOOLS

Panoramic mode calibration
Technical Manual
ADJUSTMENT & CALIBRATION
Alignment ruler (part number 50973). For checking the position of the patient positioning
mechanism.
Planmeca Proline EC X-ray unit D-1