Viavi ATC-5000NG Operation Manual page 257

Atc/dme test set
Table of Contents

Advertisement

5.7.4.16.4
BASIC/LONG ADS-B MESSAGE FEC PARITY
Command Syntax:
Description:
Numeric:
Example:
5.7.4.16.5
GROUND UPLINK REED-SOLOMON BLOCK PAYLOAD
Command Syntax:
Description:
Numeric:
Example:
5.7.4.16.6
GROUND UPLINK REED-SOLOMON BLOCK FEC PARITY
Command Syntax:
Description:
Numeric:
Example:
5.7.4.16.7
MODE STATUS PAYLOAD MESSAGE ELEMENT
Command Syntax:
Description:
Numeric:
Example:
1-2-5
Page 94
{:ATC |:ATC5000NG}{:SCE |:SCENARIO}{:STAT: |:STATIC:}<intruder
no>{:PLFEC}SP<numeric>CR
This command sets the FEC parity of the specified intruder. This command applies to ADS-B
Messages with payload type code '11' (Basic ADS-B) and '12' (Long ADS-B). The data message
is padded on the right with zeros for a length according to the payload type code (24 hexadecimal
ASCII for Basic ADS-B and 28 hexadecimal ASCII for Long ADS-B). The <intruder no> specifies
the number of the intruder in the channel grouping specified.
Basic ADS-B: 0 to FFFFFFFFFFFFFFFFFFFFFFFF (24 hexadecimal ASCII)
Long ADS-B: 0 to FFFFFFFFFFFFFFFFFFFFFFFFFFFF (28 hexadecimal ASCII)
:ATC:SCE:STAT:1:PLFEC 0\r
{:ATC |:ATC5000NG}{:SCE |:SCENARIO}{:STAT: |:STATIC:}<intruder no>{:RSB:}<block
number>{:PLMSG}SP<numeric>CR
This command sets the Reed-Solomon block payload data message of the specified intruder.
This command is valid only for Payload Type Code 14 (Ground Uplink Matrix Message). The data
message is padded on the right with zeros for a length according to the payload type code (144
hexadecimal ASCII). The <intruder no> specifies the number of the intruder in the channel
grouping specified. The <block no> specifies the Reed-Solomon block number.
<block no> 1 to 6 (decimal ASCII)
<numeric> FFFFFFFF....FFFFFF (144 hexadecimal ASCII)
:ATC:SCE:STAT:1:RSB:1:PLMSG F\r
{:ATC |:ATC5000NG}{:SCE |:SCENARIO}{:STAT: |:STATIC:}<intruder no>{:RSB:}<block
number>{:PLFEC }SP<numeric>CR
This command sets the Reed-Solomon block FEC parity of the specified intruder. This command
is valid only for Payload Type Code 14 (Ground Uplink Matrix Message). The data message is
padded on the right with zeros for a length according to the payload type code (40 hexadecimal
ASCII). The <intruder no> specifies the number of the intruder in the channel grouping specified.
The <block no> specifies the Reed-Solomon block number.
<block no> 1 to 6 (decimal ASCII)
<numeric> FFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFFF (40 hexadecimal ASCII)
:ATC:SCE:STAT:1:RSB:2: PLFEC FF\r
{:ATC |:ATC5000NG}{:SCE |:SCENARIO}{:STAT: |:STATIC:}<intruder
no>{:PLMS}SP<numeric>CR
This command sets the mode status payload message element of the specified intruder. This
command applies to ADS-B Messages with payload type code of '1' and '3.' The data message
is padded on the right with zeros for a length 24 hexadecimal ASCII. The <intruder no> specifies
the number of the intruder in the channel grouping specified.
0 to FFFFFFFFFFFFFFFFFFFFFFFF (24 hexadecimal ASCII)
:ATC:SCE:STAT:1:PLMS AF\r
ATC-5000NG ATC/DME Test Set Operation Manual
Subject to Export Control. See Cover Page for details.
Operation, Remote Operation
Scenario Commands
Rev. L0
February 2020

Advertisement

Table of Contents
loading

This manual is also suitable for:

138156