Interrogation Frequency; Interrogation On/Off; Interrogation Test Type; Interrogation Top Antenna Power - Viavi ATC-5000NG Operation Manual

Atc/dme test set
Table of Contents

Advertisement

5.9.4

INTERROGATION FREQUENCY

Command Syntax:
Description:
Numeric:
Default:
Example:
Query:
Return:
5.9.5

INTERROGATION ON/OFF

Command Syntax:
Description:
Example:
Query:
Return:
5.9.6

INTERROGATION TEST TYPE

Command Syntax:
Description:
Numeric:
Default:
Example:
Query:
Return:
5.9.7

INTERROGATION TOP ANTENNA POWER

Command Syntax:
Description:
Numeric:
Default:
Example:
Query:
Return:
1-2-5
Page 106
{:ATC |:ATC5000NG}{:XPDR}{:FREQ |:FREQUENCY}SP<numeric>CR
This command sets the transmission frequency of six generators.
952 to 1223 MHz (decimal ASCII)
1030 MHz
:ATC:XPDR:FREQ 1031.01\r
:ATC:XPDR:FREQ?\r
1031.01
{:ATC |:ATC5000NG}{:XPDR}{:RF}SP{ON |OFF}CR
This command turns on or off the interrogations. This command is equivalent to start test.
:ATC:XPDR:RF ON\r
:ATC:XPDR:RF?\r
ON
{:ATC |:ATC5000NG}{:XPDR}{:TYPE}SP<numeric>CR
This command sets the transponder interrogation test type. This command must be sent before
to define the pulse parameters.
0 to 3 (decimal ASCII)
Value
0
1
2
3
4
0
:ATC:XPDR:TYPE 1\r
:ATC:XPDR:TYPE?\r
1
{:ATC |:ATC5000NG}{:XPDR}{:POW |:POWER}SP<numeric>CR
This command sets the antenna power for the top antenna. The bottom antenna power uses this
value plus the antenna power deviation setting.
5 to -110 dBm (decimal ASCII)
-20 dBm
:ATC:XPDR:POW -21\r
:ATC:XPDR:POW?\r
-21.0
ATC-5000NG ATC/DME Test Set Operation Manual
Subject to Export Control. See Cover Page for details.
Operation, Remote Operation
Transponder Commands
Mode
Single Interrogation
Double Interrogation
Interrogation Table (Multiple)
Block Transmission
Interrogation with CW
Rev. L0
February 2020

Advertisement

Table of Contents
loading

This manual is also suitable for:

138156

Table of Contents