ATC-5000NG PERFORMANCE SPECIFICATIONS
NOTE
A 60 minute (1 hour) warm-up period is required for all specifications.
Specifications are subject to change without notice.
Please refer to these notes which are referenced throughout product specifications.
1
Absolute output power under any test condition is limited to that shown under "Power" in the
"TRANSMITTER" specification section.
2
Pulse Width/Position Exception: Block Mode timing accuracy +/-20 ns.
3
Any test configuration which results in SPR to P5 falling edge spacing >0.5 µs is not permitted
and may result in unpredictable P5 behavior.
4
All-Call Long interrogations are limited to a PRF range of 1 Hz to 1 kHz.
5
Pulse-timing, measurement-accuracy specifications applicable for input signals at amplitudes
+30 dBm.
Rev. L0
February 2020

Section 3 - Specifications

ATC-5000NG ATC/DME Test Set Operation Manual
Subject to Export Control. See Cover Page for details.
Specifications, ATC-5000NG Performance Specifications
,
1-3-1