Turb Measurement Settings; Settings For Turbidity Sensors - wtw inoLab Multi 9310 IDS Operating Manual

Digital meter for digital ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9310 IDS
Setting menu of the
®
VisoTurb
900-P
ba75937e10
08/2021
Menu item
Temp. comp. (TC) /
Linear coeff.
Temp. comp. (TC) /
Reference temp.
Multiplier for TDS

10.5 Turb measurement settings

10.5.1 Settings for turbidity sensors

The settings are made in the menu for the measured parameter, turbidity. To
open the settings, display the required measured parameter in the measured
value display and press the
switch to the measured value display with
The possible settings are individually displayed for each sensor. Default set-
tings are printed in bold.
Menu item
Calibration /
Calibration record
Calibration /
Calibration data
storage / Display
Calibration /
Calibration data
storage / Copy to
USB flash drive
Calibration /
Calibration data
storage / Output to
RS232/USB
Possible
Explanation
setting
Coefficient of the linear tempera-
0.000 ...
ture compensation.
2.000 ...
This menu item is only available
3.000 %/K
when the linear temperature com-
pensation is set.
Reference temperature
20 °C
This setting is only available for the
25 °C
measured parameters, conductivity
(ϰ) and resistivity (ρ).
Multiplier for TDS value
0.40 ... 1.00
<ENTER>
key. After completing the settings,
<M>
Possible
Explanation
setting
Displays the calibration record of the
-
last calibration.
Displays the last calibration records
-
(max. 10)
Outputs the stored calibration data
-
to a connected USB memory device/
USB printer)
Outputs the calibration data storage
-
to the interface
.
Settings
77