5.2.9 QSC function (sensor quality control)

The QSC function (Quality Sensor Control) is a new sensor evaluation for
digital IDS sensors. It evaluates the condition of an IDS pH sensor individually
and with a very fine grading.
The QSC scale shows the current sensor evaluation with an indicator on the
display.
0
pH
7.007
01.02.2014
i
08:00
1 QSC scale
The double arrow on the QSC scale indicates the current sensor
evaluation
In the USB output the sensor evaluation is given as a percentage (1-100).
The finely graded sensor evaluation of the QSC function promptly calls your
attention to changes of the sensor.
Thus you can do what is necessary to restore the optimum measuring quality
(e.g. clean, calibrate or replace the sensor). clean, calibrate or replace the
sensor).
With QSC function
Very fine grading of the sensor eval-
uation (100 grades)
The reference value is individually
determined for each sensor during
the QSC initial calibration.
Low tolerances for zero point and
slope when using QSC buffer solu-
tions
Additional QSC calibration required
(with special QSC buffer set)
The QSC function is enabled by once carrying out an additional three-point cali-
bration with special QSC buffer solutions. It covers the measuring range of the
sensor (pH 2 to pH 11). The QSC initial calibration determines the actual condi-
tion of the sensor and stores it as a reference in the sensor.
To meet the high requirements of a QSC initial calibration, the QSC initial cali-
bration should optimally be carried out with the initial commissioning of the
sensor.
14
25.0
°C
USB output
Without QSC function (sensor sym-
bol)
Rough grading of the sensor evalua-
tion (4 grades)
A theoretical reference value is used
for all sensors
Greater tolerances for zero point and
slope when using commercial buffer
sets
No additional calibration required
®