Calibration Points; Calibration Data - wtw inoLab Multi 9310 IDS Operating Manual

Digital meter for digital ids sensors
Table of Contents

Advertisement

pH value
36

5.2.6 Calibration points

Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:
Calibration
1-point
2-point
3-point to 5-
point
You can display the slope in the units, mV/pH or % (see section
10.1.1 S
ETTINGS FOR P

5.2.7 Calibration data

The calibration record is automatically transmitted to the interface
after calibrating.
The calibration record of the last calibration is available in the menu Calibration
/ Calibration record. To open it in the measured value display, press the
<CAL__>
key.
The calibration records of the last 10 calibrations are available in the menu Cali-
bration / Calibration data storage. To open the Calibration menu in the
measured value display, press the
Determined
Displayed calibration data
values
 Zero point = Asymmetry
Asymmetry
 Slope = Nernst slope
 Zero point = Asymmetry
Asymmetry
Slope
 Slope = Slope
 Zero point = Asymmetry
Asymmetry
Slope
 Slope = Slope
The calibration line is calculated by
linear regression.
H
MEASUREMENTS
<ENTER>
®
inoLab
Multi 9310 IDS
(-59.2 mV/pH at 25 °C)
, page 69).
key.