Menu item
Calibration /
Calibration data
storage / Output to
RS232/USB
Calibration /
Calibration interval
Calibration /
Comparison meas.
FDO Check /
Start FDO Check
®
(only for FDO
925)
FDO Check / Check
interval
®
(only for FDO
925)
Sal correction (only
for parameter mg/l)
Salinity
(only for parameter
mg/l)
Response time t90
Reset
Possible
Explanation
setting
Outputs the calibration data stor-
-
age to the interface
Calibration interval for the D.O.
1 ... 180 ...
sensor (in days).
999 d
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring screen.
Enables to adjust the measured
on
value with the aid of a comparison
off
measurement, e.g. Winkler titra-
tion.
For details, see section 7.3 C
BRATION
Starts the check with the
-
®
FDO
Interval for the FDO Check (in
1 ... 60 ...
days).
999 d
The meter reminds you to check
the sensor regularly by FDO Check
status indicator in the measuring
screen.
Manual salt content correction for
on
concentration measurements.
off
Salinity or salinity equivalent for the
0.0 ... 70.0
salt content correction.
This menu item is only available if
the manual salinity correction is
switched on.
Response time of the signal filter (in
30 ... 300 s
seconds).
A signal filter in the sensor reduces
the limits of variation of the mea-
sured value. The signal filter is
characterized by the response time
t90. This is the time after which 90
% of a signal change is displayed.
Resets all sensor settings to the
-
delivery condition (see section
10.7.1 R
MENT SETTINGS
®
inoLab
Multi 9310 IDS
, page 50.
Check
ESETTING THE MEASURE
, page 81).