Calibration Data - wtw inoLab Multi 9310 IDS Operating Manual

Digital meter for digital ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9310 IDS
Continuing with
three-point calibra-
tion
Displaying the cali-
bration data
ba75937e10
08/2021
TRB
124.0
1010.0 FNU
Cal
01.07.2011
16.
If necessary, terminate the calibration as a two-point calibration with
<M>
.
The new calibration values are displayed.
or
Continue with three-point calibration.
Repeat the steps 11 to 15 with the third standard solution. The new calibration
values are displayed after the last calibration step was completed.

9.2.5 Calibration data

Calibration The calibration record of the last calibration is available in the menu
<ENTER>
/ Calibration record. To open it in the measured value display, press
<CAL__>
the
key.
The calibration records of the last 10 calibrations are available in the menu Cali-
bration / Calibration data storage / Display. To open the Calibration menu in the
measured value display, press the
Turbidity measurement (VisoTurb
3
Standard
FNU
08:00
<ENTER>
key.
®
900-P)
67

Advertisement

Table of Contents
loading

This manual is also suitable for:

Inolab multi 9310p ids

Table of Contents