Measuring The Temperature; Orp Calibration - wtw inoLab Multi 9310 IDS Operating Manual

Digital meter for digital ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9310 IDS
ba75937e10
08/2021
ture of the test sample.
When a sensor with integrated temperature sensor is operated, the tempera-
ture value is displayed and recorded together with the measured value.

6.1.3 Measuring the temperature

For reproducible ORP measurements, it is essential to measure the tempera-
ture of the test sample.
When operating a sensor without integrated temperature sensor, you first have
to measure and enter the temperature of the sample.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
Temperature
Resolution of the
sensor
temp. display
yes
0.1 °C
-
1 °C
6.2

ORP calibration

ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.
Temp. measurement
Automatic with temperature
sensor
Manual
ORP
45

Advertisement

Table of Contents
loading

This manual is also suitable for:

Inolab multi 9310p ids

Table of Contents