Settings
60
11
Settings
11.1 Measurement settings
11.1.1 Settings for pH measurements
The settings for pH measurements are made in the menu for calibration and
measurement settings of the pH/ORP measurement. To open the settings, dis-
play the required measured parameter in the measured value display and
<OK>
press the
key. After completing the settings, switch to the measured
<MODE>
value display with
In the following table, only those settings are listed that concern the pH mea-
surement.
Default settings are printed in bold.
Menu item
Calibration /
Calibration record
Calibration /
Calibration data
storage / Display
Calibration /
Calibration data
storage / Output to
USB
Calibration / Buffer
Calibration / One
point calibration
.
Possible set-
Explanation
ting
Displays the calibration record of
-
the last calibration.
Displays the calibration record.
-
Further options:
Scroll through the calibration
records with
Output the displayed calibration
record to the interface with
<F2>
Output all calibration records to
the interface with
output].
Quit the display with
[Back] or
Switch directly to the measured
value display with
Outputs the calibration records to
-
the interface.
Buffer sets to be used for pH cali-
TEC
bration (see section 6.2 C
AnyCal
NIST/DIN
TION
TEC 2
...
Quick calibration with 1 buffer
Yes
No
MU 6100 H
<
><
>
/[USB output].
<F2__>
<F1>
<OK>
.
<MODE>
, page 29).
ba77077e06
.
/[USB
/
.
-
ALIBRA
05/2021