Oxi 3310 IDS (Oxi 3315)

Settings

ba75944e02
10/2014
6
Settings
6.1
D.O. measurement settings
6.1.1 Settings for D.O. sensors
(menu for measurement and calibration settings)
The settings are available in the menu for measurement and calibration set-
tings. To open the settings, display the required measured parameter in the
measured value display and press the <ENTER> key. After completing the set-
tings, switch to the measured value display with <M>.
Default settings are printed in bold.
Menu item
Calibration /
Calibration record
Calibration /
Calibration data
storage / Display
Calibration /
Calibration data
storage / Output to
RS232/USB
Calibration /
Calibration interval
Calibration /
Comparison meas.
FDO Check /
Start FDO Check
FDO Check / Check
interval
Sal correction
Possible
Explanation
setting
-
Displays the calibration record of the
last calibration.
-
Displays the last calibration records
(max. 10)
-
Outputs the calibration data memory
to the interface
1 ... 180 ...
Calibration interval for the D.O. sen-
999 d
sor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
on
Enables to adjust the measured
off
value with the aid of a comparison
measurement, e.g. Winkler titration.
For details, see section 5.3 C
, page 28.
TION
-
Starts the FDO
1 ... 60 ...
Interval for the FDO Check (in days).
999 d
The meter reminds you to check the
sensor regularly by FDO Check sta-
tus indicator in the measuring
screen.
on
Manual salt content correction for
off
concentration measurements.
Settings
-
ALIBRA
®
check procedure
33