After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
For evaluation, the characteristic curve of the sensor that was
determined is compared to the characteristic curve of an ideal sen-
sor at similar environmental conditions (relative slope S): S = S
/ S
sor
ideal sensor
Display
Error
Error elimination (see section 15 W
page 98)
Setting/
Explanation
function
-
Displays the calibration records.
Further options:
 Scroll through the calibration
 Output the displayed calibra-
 Output all calibration records to
 Quit the display with
 Switch directly to the mea-
-
Outputs the calibration data stor-
age to the interface (see section
12 T
120).
. An ideal sensor has a slope of 1.
Calibration record
+++
++
+
Error
HAT TO DO IF
Dissolved oxygen
<
><
records with
tion record to the interface with
<F2>
/[USB output].
the interface with
<F2__>
[USB output].
<F1>
<ENTER>
[Back] or
.
sured value display with
, page
RANSMITTING DATA
Relative slope
S = 0.94 ... 1.06
S = 0.92 ... 0.94
or
S = 1.06 ... 1.08
S = 0.90 ... 0.92
or
S = 1.08 ... 1.10
S < 0.90
or
...,
S > 1.10
>
.
/