What to do if...
Display
Meter does not react
to keystroke
You want to know
which software
version is in the
meter or IDS sensor
102
Cause
– Batteries almost empty
Cause
– Operating condition undefined
or EMC load unallowed
Cause
– E. g., a question by the service
department
inoLab
Remedy
– Replace the batteries (see section
14.1 M
AINTENANCE
Remedy
– Processor reset:
<ENTER>
Press the
key simultaneously
Remedy
– Switch on the meter.
<ENTER__>
– Open the menu,
age & config / System / Service infor-
mation. The instrument data are
displayed.
or
– Connect the sensor.
<F1>
Press softkey [
[More] The sensor data are displayed
(see section 4.1.5 S
18)
ba75937e10
®
Multi 9310 IDS
, page 96)
<On/Off>
and
/ Stor-
<F1>
] /
Info
, page
ENSOR INFO
08/2021