wtw inoLab Multi 9310 IDS Operating Manual page 102

Digital meter for digital ids sensors
Table of Contents

Advertisement

What to do if...
Display
Meter does not react
to keystroke
You want to know
which software
version is in the
meter or IDS sensor
102
Cause
– Batteries almost empty
Cause
– Operating condition undefined
or EMC load unallowed
Cause
– E. g., a question by the service
department
inoLab
Remedy
– Replace the batteries (see section
14.1 M
AINTENANCE
Remedy
– Processor reset:
<ENTER>
Press the
key simultaneously
Remedy
– Switch on the meter.
<ENTER__>
– Open the menu,
age & config / System / Service infor-
mation. The instrument data are
displayed.
or
– Connect the sensor.
<F1>
Press softkey [
[More] The sensor data are displayed
(see section 4.1.5 S
18)
ba75937e10
®
Multi 9310 IDS
, page 96)
<On/Off>
and
/ Stor-
<F1>
] /
Info
, page
ENSOR INFO
08/2021